| Measuring method |
Right angle scattered light method |
| Measuring range |
0.000 to 100.0 NTU |
| Output range |
Configurable within the measuring range
Minimum range: 0 to 0.2 NTU
Maximum range: 0 to 100 NTU
Minimum span: 20 % or more of upper limit of the range or 0.2 NTU, whichever is greater. |
| Analog output |
2 outputs:
output 1: 4 to 20 mA DC, isolated
output 2: 4 to 20 mA DC or 0 to 20 mA DC, selectable, isolated (Maximum load resistance 550 Ω)
|
| Serial communication |
1 output, RS-422 or RS-232C, isolated |
| Contact output |
3 contacts |
| Contact input |
2 contacts |
| Calibration |
Zero/span calibration |
| Ambient temperature |
-5 to 50 deg C (Sample and tap water may need protection against freezing) |
| Ambient humidity |
5 to 95 % RH (non-condensing) |
| Sample water conditions |
Flow rate: 0.05 to 20 l/min
Temperature: 0 to 50 deg C
Pressure: 500 kPa max. |
| Power supply |
100 to 240 V AC -15 %/+10 %, 50/60 Hz |
| Power consumption |
Converter + Detector: 50 VA max. |