I’m measuring temperature, but the recorder's temperature display is higher than the actual temperature. Can I correct the temperature on the recorder? I don't have the broken line correction option (/CC1).

Yes, by setting the range rating mode to Scale.

For example, with a K thermocouple and measured values with an input span of 0.0°C–100.0°C, to lower all temperatures by 1°C, enter the following settings.

Mode: Scale
Range: TC
Type: K
Span lower limit: 1.0
Span upper limit: 101.0
Scale lower limit: 0.0
Scale upper limit: 100.0
Unit: °C

Súvisiace produkty a riešenia

  • Button Operated DX1000/DX2000

    The Daqstation DX1000/2000 series is a data acquisition system with display. Its HMI supports custom graphics for industry applications, audit trail and advanced security conforming to FDA 21CFR Part 11.

    Vidieť viac

Hore