The following statuses and sub-statuses are covered by Wireless Input Unit Error:
- Status CAUTION, sub-status LOW_BATTERY: Battery abnormality.
*The battery of the GX70SM is low - Status CAUTION, sub status DEAD_BATTERY: Battery abnormality
*The battery of the GX70SM is dead - Status CAUTION, sub status UNIT_CHANGE: Device change occurred
*Only occurs with the /AS option (the advanced security function)}
*A GX70SM with a different serial number from the reconstructed GX70SM is connected - Status ERROR, sub-status ERROR: Malfunction
*Malfunction of the GX70SM *1
*1 :Malfunction of the GX70SM refers to the following conditions:- Abnormal settings and calibration values
- A/D abnormality
- Hardware, CPU abnormality
- Flash/FRAM sum abnormality
- Mode setting abnormality
- Burnout
- Status FAILED, sub-status: Communication lost
*Communication with the GX70SM has not occurred for about 90 seconds