I set up time setting on the µR10000/µR20000 remote control function. I input a contact, but the time does not set. What should I check?

The time window in which you can input the contact and then set the time is limited.
While the time on the µR is between 2 m 00 s and 57 m 59 s, the time is not changed. Input the contact while the time on the µR is between 58 m 00 s and 1 m 59 s.

Zugehörige Produkte & Lösungen

  • µR10000 & µR20000

    Die μR-Serie der kompakten industriellen Papierschreiber mit Aufzeichnungsbreiten von 100 mm (μR10000) und 180 mm (μR20000).
     


Top