Measuring Water Quality at Water Purification Plants

Introduction

Improving potable water quality, safety, and sustainability of the drinking water supply are our mission for the future. To achieve them, measuring water quality accurately and contentiously are very important. The measurement of turbidity in potable water facility and distribution line is a vital component of ensuring safe and clean drinking water.
FLXA402T/TB820D right angle scattered light turbidity analyzer is used to control and monitoring drinking water quality. WHO says in "Guidelines for drinking-water quality" that "to ensure effectiveness of disinfection, turbidity should be no more than 1NTU and preferably much lower. Large, well-run municipal supplies should be able to achieve less than 0.5 NTU before disinfection at all times and should be able to average 0.2 NTU or less." FLXA402T/TB820D can measure from 0-2 NTU to 0-500 NTU with excellent accuracy and have been contributed to healthy life. To achieve continuous measurement, FLXA402T/TB820D equipped useful functions, for example diagnostics, automatic cleaning and automatic zero calibration. These functions help to improve water quality control.

 

Expected Benefits

  • Measures low turbidity levels both continuously and accurately
  • Reduces operating cost
  • Eliminates the need for manual cleaning
  • Keeps the initial cost of equipment replacement to a minimum
  • Multiple sensor input to a converter is available (Turbidity, Residual chlorine, pH and Conductivity)

Process
Process

 

Process Overview

The potable tap water distributed to individual households is originally produced from raw water taken from rivers or underground springs. At water purification plants, chemicals are injected into the raw water to produce flocs that serve to absorb hazardous substances in the raw water and allow them to be deposited and filtered out. Chlorine is then added as a disinfectant to the treated water. A variety of water quality measuring instruments, including turbidity analyzers, residual chlorine analyzers, and pH meters, are used in water purification plants. This application note focuses on turbidity analyzers, which are used to verify that filtration systems are operating in excellent condition. The FLXA402T/TB820D turbidity analyzer continuously measures post-filtration turbidity in order to achieve the required turbidity control value of 0.1 mg/L.

 

Solution Details

Measurement system

Right angle scattered light turbidimeter
Model: FLXA402T-A-B-□□-TB-□□-N□-WR-N-□-N-NN-NN
TB820D-□□-□□-□□-NN-CT-NN/D□
Main components: Turbidity detector, converter, pressurized head tank
Primary product specifications:
Measuring range: 0-0.2 to 0-500 NTU
0-0.2 to 0-700 FNU/NTU
Output signal: 4-20 mA DC, Modbus TP/IP, Modbus RTU (RS-485)
Installation location: indoors
Zero-turbidity filters (to be purchased separately)
Filter assembly 1 micron: P/N K9008ZD
Filter assembly 0.2 micron: P/N K9726EH

 

Utilities for FLXA402T/TB820D

Power supply: 100-240 V AC, 50/60 Hz Power
Power consumption: Approximately 120 VA

Note on installation

This turbidity analyzer uses a pressurized head tank to eliminate air bubbles, which are a major error factor in the measurement of low turbidity levels.

Utilities for FLXA402T/TB820D

Utilities for FLXA402T/TB820D

Industrien

  • Water & Wastewater

    Water resources are finite, and therefore contributing to a sustainable water cycle is one of the Sustainable Development Goals (SDGs). Yokogawa has been providing advanced digital control solutions for the stable supply of clean and safe water, wastewater treatment for protecting the water environment, water loss management and optimization of plant operation for reducing CO2 emissions and running costs. With our leading-edge technologies, dependable products and extensive expertise and experience of diverse water projects around the world, we work with you to provide sustainable water solutions that boost your business and add value throughout the plant lifecycle.

    Yokogawa supports a wide range of water control applications in both the municipal and industrial water markets.

      

     

    Mehr anzeigen
  • Water Treatment

    Yokogawa provides optimum solutions to improve the reliable operation of water treatment plants. These solutions include sophisticated technologies, unsurpassed product quality, and extensive application know-how throughout the lifecycle.

    Mehr anzeigen

Zugehörige Produkte & Lösungen

  • Trübungsdetektor TB820D mit 90°-Streulichtwinkel

    Der TB820D ist ein digitaler Trübungsanalysator der nächsten Generation und kann an den FLXA402T angeschlossen werden. Der Trübungsanalysator TB820D ist mit einer LED als Lichtquelle, einer flachen Fensterzellstruktur, neuen Diagnosefunktionen, einer Ultraschall-Reinigungsfunktion usw. ausgestattet. Diese tragen zu einer maßgeblichen Reduzierung der nötigen Wartungszeit bei.

    Mehr anzeigen
  • Trübungsanalysatoren

    Der TB820D ist ein digitaler Trübungsanalysator der nächsten Generation und kann an den FLXA402T angeschlossen werden. Der Trübungsanalysator TB820D ist mit einer LED als Lichtquelle, einer flachen Fensterzellstruktur, neuen Diagnosefunktionen, einer Ultraschall-Reinigungsfunktion usw. ausgestattet. Diese tragen zu einer maßgeblichen Reduzierung der nötigen Wartungszeit bei.

    Mehr anzeigen

Top