Does the PH21 have a liquid Earth?

No, the PH21 does not have a Liquid Earth (LE) electrode.
The LE is not needed in many of the applications we target for this sensor. When necessary, a separate Liquid Earth connection can be connected to the SA11 using the screw terminal on the SA11 mounting bracket.
Number 14 of the cable can be connected as normal. Provisions have been made in the connector to utilize the reference (13) for diagnostics. Only the glass breakage diagnostics is then available. When connecting a separate liquid earth, as with the pH fittings (FF20/FS20/FD20), all the diagnostics will be available.

Wide Body Sensor for Water and Wastewater PH21

Zugehörige Produkte & Lösungen

  • Digitaler SENCOM™ SMART-Adapter, SA11

    Wiederverwendbarer SMART-Adapter, bei dem nur der Analogsensor am Ende der Lebensdauer entsorgt werden muss. Mit der Plattform SENCOM 4.0 reduziert Yokogawa Kosten und Abfälle und trägt damit zu seinem langfristigen Unternehmensziel einer nachhaltigen Zukunft für alle bei.

  • Durchfluss/NPT-Armaturen FF20/FS20

    Yokogawa hat viel Aufwand und Zeit in die Entwicklung einer breiten Palette von Armaturen investiert. Ein besonderer Schwerpunkt lag dabei auf der Senkung von Installations- und Wartungskosten und somit günstigeren Betriebskosten.