Introduction
For years, many types of high-density culture methods, such as spheroids, fiber scaffolds, and extracellular matrixes, have been proposed for in vitro cell-based assays. These culture systems have been recognized to more accurately simulate a cells natural environment than standard monolayer cultures on a flat substrate. Therefore, cells in high-density culture conditions are expected to exhibit responses against chemical treatments that closely resemble responses of tissues in vivo.
General homogeneous assay protocols originally developed for monolayer culture can be applicable to these high-density culture systems with minor modifications.
However, for microscopy and other image-based assays, there are significant obstacles to overcome when applying conventional image analysis protocols to high-density cultures. The optical architectures of most microscope-coupled research instruments do not capture light based information throughout the entire thickness cell aggregates due to depth-of-field limitations. In addition, image-analysis software optimized for monolayer cell culture are not able to perform cell-by-cell object recognition and resulting quantification.
Here we show a set of examples with an ultra-high density HepG2 (hepatocellular carcinoma) cell culture to explain how the CQ1 can capture clear images from entire thickness of a cell layer three-dimensionally. Additionally, we will show how the CQ1 can analyze cell responses against chemical treatment, on a cell-by-cell basis.
The protocol on this note has a potential applications for analyses of various high-density cultures, including three dimensional cell culture conditions.
Figure 1. Images of highly dense HepG2 cultures and nuclear contouring.
Non-treated (a) or staurosporine (10-7 M for 48 h) treated (c) cells.
Three dimensional cell-by-cell nuclear recognition was carried out by using the Spheroid Analysis Algorithm of the CQ1 software (b and d). Objective lens: 20X.
Experimental procedure
•A 96-well glass bottom microplate was coated with an extra cellular matrix, Matrigel (5 fold dilution with culture medium).
•HepG2 (hepatocellular carcinoma) cells were seeded on the pre-coated plate at a density of 5X104 cells/well. The plate was incubated for 48 hours to create an over-confluent state.
•Staurosporine was added to experimental wells and the plate was incubated for and additional 48 hours.
•Cells were fixed with a formaldehyde solution then tagged with anti-active caspase-3 and anti-H3Ser10P. Bound primary antibodies were visualized with fluorescently labelled secondary antibodies. Cell nuclei were stained with Draq7 in the presence of RNaseA.
•Cell images were captured and the images analyzed by the CQ1. Graphing and statistical processing was carried out using FCS Express™ 5 Image Cytometry (De Novo Software, Glendale, CA) (optional).
Results and discussions
An ultra high-density HepG2 cell culture was created and cell-by-cell analysis was carried out by the CQ1 to evaluate cell responses against staurosporine, a hepatotoxic chemical.
•To facilitate dense cell layer formation, the plate was coated with an extracellular matrix. In preliminary tests, compared with culture on a non-coated normal plastic-bottom plate, HepG2 cells plated on the matrix-coated substrate exhibited approximately ten-fold higher sensitivity toward staurosporine toxicity (data not shown).
•Images of fluorescently labelled cells were captured in three-dimensions from the thick cell layer formation (Fig 2). Two molecular markers, H3Ser10P (cell growth) and active caspase-3 (apoptosis) were selected for immunofluorescent labelling.
•Segmentation of individual cell nuclei enabled cell-by-cell characterization in response to sturosporine (Fig 3).
•The CQ1 is a versatile system that allows simultaneous analysis of multiple markers and parameters at the single cell level in high density cell culture based assays, including thick cultures of hepatocytes.
Figure 2. 3D reconstitution of HepG2 cell images.
Twenty-one slices (along the Z-axis) of a multicolor image encompassing 50 µm of thickness were reconstructed to create a 3D image. Non-treated (a) or staurosporine treated (b) cells were fluorescently immunostained with anti-H3Ser10P (magenta) and anti-active caspase-3 (green). Cell nuclei were stained with Draq7 (gray). Objective lens: 20X.
Figure 3. Multi-parametric analysis of two cell markers
Image analysis data from the CQ1 of non-treated (a) or staurosporine treated (b) cells was exported and further analyzed by scatter plots in FCS Express™ 5 Image Cytometry. The proportions of the growing or dead cell populations were evaluated quantitatively. On each scatter plot, 1X104 events were plotted.
Our Social Medias
We post our information to the following SNSs. Please follow us.
Follow us | Share our application | |
@Yokogawa_LS | Share on Twitter | |
Yokogawa Life Science | Share on Facebook | |
Yokogawa Life Science | Share on LinkedIn |
Yokogawa's Official Social Media Account List
Zugehörige Produkte & Lösungen
-
Benchtop-System
Das CQ1 bietet Konfokalbilder und erweitertes Live-Cell-Imaging der höchsten Qualität in einem raumsparenden Benchtop-Design.
-
CellPathfinder
CellPathfinder ist für unsere HCA-Systeme, CQ1 und die CellVoyager Serien, konzipiert. Ob Sie Anfänger oder Experte sind, die Analysis-Software ermöglicht es Ihnen, subtile physiologische Veränderungen und selbst ungefärbte Samples mit unterschiedlichen Grafikoptionen zu quantifizieren.
-
CellVoyager High-Content Analysis System CQ3000
CQ3000 can acquire high-resolution 3D images at high speed while culturing cells by combining options according to the application.
-
High-Content-Screening
CellVoyager CV8000 ist das fortschrittlichste High-Content-Screening-System. Der optimierte eingebaute Inkubator gewährleistet bessere Umgebungsbedingungen für Ihre Langzeit-Live-Cell-Experimente. Mit seiner Ausbaufähigkeit zu bis zu vier Kameras, fünf Lasern und einem optionalen Pipettier-Roboter erlaubt das System die Durchführung komplexer Assays und High-Content-Screening-Experimente.
-
High-Content-Analysesystem CellVoyager
Unsere High-Content-Analyse-Systeme (HCA) verwenden eine leistungsstarke Analysesoftware, um ein möglichst breites Spektrum an Forschungsanwendungen abzudecken; von der Grundlagenforschung bis hin zum komplexen Wirkstoff-Screening.
-
Life Science
Die High-Content-Analyse-Systeme (HCA) und Dual-Spinning-Disk-Technologien von Yokogawa ermöglichen hochauflösendes Hochgeschwindigkeits-Live-Cell-Imaging für die Spitzenforschung rund um den Globus.