- Check the power supply wiring.
- Check whether the power supply voltage meets the specifications.
- Confirm that the accessory RJC sensor is connected to the input terminal.
- Short the input terminals, then confirm that the output is equivalent to the terminal temperature.
- If the output signals is of the current specifications, attach an external 250 Ω resistance and measure the voltage.