DA100-1
Mededeling: Dit product is niet meer leverbaar vanaf Mar 31, 2015. Zie dit vervangende product:

Middelen

If 300 ch or less, you can set it so that the subunit is not recognized.
Specify the certificate when ordering the product. Test certificates are included with the test certificates of the main unit and modules.
Yes. This is possible both with the RS-232C and Ethernet models. Aside from the DA100 Quick Start package, this requires the extender base (DV100-012), expansion cable (DV200-XXX), and input module. The thinking is the same as the DA100 standalone mo...
Ex. 1 Using alarm settings and an AND relay Input channel alarm settings Level1=H Alarm value=100.0 (°C) Relay (internal SW) setting = Output relay number Level2=L Alarm value=500.0 (°C) Relay (internal SW) setting=Output ...
It is a Yokogawa proprietary protocol that is HDLC compliant.
The time in the trends is the time on the DA100. When connected for a long time, the times on the PC and DA100 can start to diverge. Closing and then reopening the connection will synchronize the times because the PC's time is automatically set o...
There are 2 possible reasons for an error code 139.   The floppy disk is damaged. Use a different floppy disk (one from a different production lot). The floppy drive is damaged. Please submit for repair. ...
Double-click the DAQ32Plus environment settings icon, click the Environment tab, and make sure ETHERNET is selected and that the DA100's IP address is set. If the IP address is not set, set it and then restart DAQ32Plus.
Set the span and scale to ±30000 (ignoring the decimal point). In this case the range is 250 V, 5 A, so the span is in the range of ±1250 W. As you are dropping the current with the CT to 1/10, you have to increase the scale by 10 times...
To write the current settings to the DA100, "Send" the settings, then "Receive" them, then immediately execute "Balance." When you change settings, sometimes the "Balance" menu does not become active.
If connecting the PC directly to the DA100, use one cross cable. If going through a hub, use 2 straight cables.
Yes. If you are monitoring or recording with DAQ32, stop those. Then, change the speed as follows. Start "Environment, diagnosis and calibration", and click the Environment tab. In the RS-232C field, click 9600 under Baud Rate to displa...
Yes. The DA100 does not keep track of absolute time. When starting to acquire measured data using package software on a PC, the PC's time is set on the DA100. After that, time is managed based on the DA100's scan interval (relative time). The...
Because DAQ32 Plus has greater functionality than DAQ32, it requires that much more CPU speed and memory. If there was no problem on DAQ32 but the performance bar goes to OVER on DAQ32 Plus, probably the baud rate set on DAQ32 is too slow for DAQ32 P...
In "Shipping and storage conditions (unpowered)" of the DA100 GS, it gives a specification of "Shock: 329 m/s2 or less (in packaging)". Given a shock of 329 m/s or less while packaged, it might need readjustment, but it should ret...
This requires upgrading the DARWIN hardware style. Please submit for factory rework.
Generally they are not initialized if the modules are not different from the previous system. However, initialization happens if slots are in the following conditions. Slot replaced with a module of a different model name or style number Slot who...
You can, starting with style 8 of the DA. Setting command: VS. Control execute command: VD. The DR and DC do not support this. With DAQ32 and DAQ32Plus you can set up output relays, but you can't actually operate them with commands. You can also ...
Is there a limit on the extender modules?
(ns-faq-darwin-8052-select)
Extender modules can be connected to the main unit (DA100-1) and subunit (DS400, DS600). However, only 1 extender module can be connected to each. Total cable length: expansion cable (max. 500 m) + extender cable (max. 30 m). Even if using an extende...
Yes. The integral value is the scale calculation.
There is no strict specification for the input current value of DCV/TC INPUT, but the current value can be determined from the following specification. DCV/TC INPUT specification Max input voltage: V 2 VDC range, thermocouple, RTD: ±10...
No. You can use an STP cable of CAT5 or higher. The maximum distance between units is 100 m. You can use either a straight or cross cable.
For 0.5 second models, temperature is measured for RJC every 2.5 seconds (with a 0.5 second scan, since the time for processing temperature measurement, burnout, and other factors is too short, measurement occurs every 5 cycles). For 2.0 second mode...
The GM10 does not support GP-IB communications. It comes standard with an Ethernet communication interface, so we recommend connecting to the PC via LAN.
On the RS232C communication module, set communication parameters by using the dip switches on the front of the module. Make sure the communication module’s dip switches match what is displayed on the DAQ32's “Environment, diagnosis, a...
The range of the strain input module depends on the measurement mode and gauge method. (see the table below) Meausrement mode 2K 20K 200K 1 gauge method -2000 to +2000µε -20000 to +20000µε...
Unless you set the counter to go back to zero with the SETUP function, an overflow results.
Overzicht:

Unless you set the counter to go back to zero with the SETUP function, an overflow results.

Thermocouple data is generally not included with our digital devices. The accuracy of data measured with thermocouples is the combined accuracies of: the thermocouple itself; the temperature compensator; the linearization that approximates the thermo...
  RELATIVE: Count begins from the moment the timer time is set, and timeup occurs when a set time is reached. For example, if timer 1 is currently at 7 days/10 hours/00 minutes, and you set 01 days/10 hours/00 minutes, the timeup occurs every...
4.2 µA flows from a burnout power supply, resistance is measured, and the determination is made. : Thermocouple is normal: Less than 10 KΩ Thermocouple breaks, and while it is breaking: 10 KΩ or more
After entering Setup mode settings (including burnout), you have to store the settings using the XE command. If you don't send the XE command, the settings are ignored.
If you specify unidentified channels, no errors occur but the channels are ignored.
When applying a moving average, the decimal places matches the number of digits in the range.
The voltage generated when a burnout current of 4.2 mA flows to the measuring terminal circuit is measured in the 60 mV range, and a determination is made. If 42 mV (10 kΩ) or more, the result is 0 (open). If less than that, 1 (closed).
The input types are non-voltage contact and open collector. Hence, it is not necessary to input a voltage to pulse input. This specification shows the withstand voltage and current capacity determined for contacts or transducers. In other words, a co...
  When measuring strain with initial balancing (EXEC), measurement is performed using the state with the gauge connected as a reference criterion. Therefore, you must balance the bridge with the gauge connected. This balancing is called initi...
No, on DARWIN, only a vertical parity check is performed.
Yes. Set the IP address, subnet mask, and default gateway on the main unit.
For voltage measurement, if 5% of range + range span is exceeded, an overrange occurs (±32767 is output). For temperature range, the +10°C measuring range is the upper limit. Ex) For the ±50 V range (span of -50 to +50), the upper l...
Does DARWIN support Windows XP?
(ns-faq-darwin-8017-spec)
R10.03 and later support Windows XP.
DAQ32 (DP120) or DAQWORX DAQ32 Plus (WX102) installed on Windows 7 or Windows Vista cannot change the IP address of DARWIN series Ethernet modules (suffix code /C7) or DARWIN series standalone models with the Ethernet communications option (model cod...
In principle, they should continue to run even if their upstream subunit has a power failure. However, temporal effects will be experienced (data acquisition will be temporarily lost, or certain data might become damaged).
The Viewer cannot determine the numbers after the decimal with statistical computation on DI input. Data acquired and results computed by the BFD32 are affected by the decimal of the specified measurement range. For DI (contact input), a 4.2 µA...
If you have not changed any settings, check the following. Check the amount of free space on the hard disk, and perform error checking Check the DA100 power, and whether the green STATUS lamp is lit Communication cable connection and replacement...

Downloads

Op zoek naar meer informatie over onze mensen, technologie en oplossingen?

Contact

Top