No, the PH21 does not have a Liquid Earth (LE) electrode.
The LE is not needed in many of the applications we target for this sensor. When necessary, a separate Liquid Earth connection can be connected to the SA11 using the screw terminal on the SA11 mounting bracket.
Number 14 of the cable can be connected as normal. Provisions have been made in the connector to utilize the reference (13) for diagnostics. Only the glass breakage diagnostics is then available. When connecting a separate liquid earth, as with the pH fittings (FF20/FS20/FD20), all the diagnostics will be available.

Related Products & Solutions
-
Digital SMART SENCOM™ Adapter, SA11
Reusable SMART adapter, requiring only the analog sensor to be disposed of when it reaches the end of its lifetime. With the SENCOM 4.0 platform, Yokogawa delivers reduced costs and waste while contributing to its long-term business goals of a sustainable future for all.
-
Flow/NPT Fittings FF20/FS20
FF20/FS20 Fittings for pH and ORP