Per the specifications, the accuracy of type K at 0°C or higher is ±0.1% ±1 digit. The measuring range in this case is that of the selected input type, -270.0 to 1000.0. You can calculate the accuracy as follows.
±0.1% of 1270°C = 1.27°C, plus 0.1°C of 1 digit = 1.37°C.
Related Products & Solutions
-
Mid-level UT55A/UT52A
Increase monitoring and operating capabilities
-
UT35A/UT32A
Increase monitoring and operating capabilities