A Gas Chromatograph analyzes a gas or volatile liquid sample by separating components for detection. The discrete separation and positive identification of components and measurement of the composition with no interference is a positive advantage of the gas chromatograph. Making good use of this advantage, the process gas chromatograph is widely used in many industries.

Yokogawa manufactures two models of the GC1000 Process Gas Chromatograph: the GC1000 S/E/C(isothermal oven type) and the GC1000 D/T/W(programmed temperature type).

The GC1000 S/E/C can analyze gas and liquid samples from room temperature up to elevated temperatures corresponding to their boiling points. The GC 1000 D/T/W is capable of performing high resolution analysis by using capillary columns similar to laboratory gas chromatograph.

   Bulletin 

   General Specification 
* Capabilities of analyzing PIONA and the Distillation points
* Capability of analyzing high boiling point liquid samples
* Capability of analyzing samples in a very wide range of boiling points
* Easy operation via User Friendly Displays
* Enhanced maintenance using PC operation
* Achievement of High Sensitivity TCD
* Expectation of the effective result by EPC (Electronic Pressure Control)
Measurable object Gas or liquid
Measurement principle Component separation by elution
Detection TCD, FID, FPD
Measurable ranges TCD 1ppm to 100%
FID 1ppm to 100%
FPD 1ppm to 0.1%
Number of streams to be measured Maximum of 31 (including standard sample)
Number of components to be measured Maximum of 255
Analysis period Maximum of 99999.9 sec
Quantifying method Absolute calibration
Sensitivity-corrected absolute calibration
Corrected area normalization
Material of the sample-contact parts Stainless steel SS316, polytetrafluoroethylene (PTFE), fluorocarbon rubber, and glass
Repeatability 1% of full scale (2σ)
* The value may vary depending on the specifications and conditions. For details, contact Yokogawa.
Analyzer specifications
Area classification TIIS Ex pd IIB+H2, T1 to T4 X
FM/CSA: X-purging, Y-purging ClassI, Div1, Groups B, C and D, T1 to T4
ATEX (CENELEC) II2G EEx pd IIB+H2, T1 to T4
Construction Drip-proof and dust-proof construction
(NEMA 3R, equivalent to IP53)
Display Liquid crystal display (LCD)
Operating ambient conditions -10 to 50 deg C, 95% RH or less
Weight Approx. 120 kg (with analyzer base sampling system)
GC1000S/E/W (Isothermal Oven)
Volume 40L (27L : with programmed Temp.Oven)
Temperature setting At fixed set point
Setting temperature range 55 to 225 deg C (Temperature can be set in one deg C steps.)
Temperature control accuracy ± 0.03 deg C
Temperature control PID control
GC1000D/T/C (Programmed Temperature Oven)
Volume 8.6L
Temperature setting Fixed point or programmed setting
Temperature setting range 60 to 320 deg C without cooler
5 to 320 deg C with cooler
Heating rate 1 to 30 deg C / min (Temperature can be set in one deg C steps.)
Temperature stability ± 0.03 deg C at a control point in the steady state.
Stability to ambient Temperatures ± 0.1 deg C / 10 deg C
Stability to supply voltage variation 0.03 deg C / 10%
Temperature control PID control
Temperature sensor Pt100 ohm RTD with over-heating prevention function.

Utility
Power supply 100 to 120V AC ± 10%, 50 / 60 Hz ± 5% (for GC1000S/E/W), or 200 to 250V AC, 50 / 60Hz ± 5% (for GC1000D/T/C)
Power consumption Maximum of 3.1kVA (for GC1000D/T/C)
Maximum of 1.5kVA (for GC1000S/E/W)
(depend on the specifications)
Instrument air
Standard (without cooler)
Pressure: 350 to 900kPa
Flow rate: 150 L/min or more (for GC1000D/T/C)
  100 L/min or more (for GC1000S/E/W)
Dew point: -20 deg C or less (depend on the specifications)
With cooler
Pressure:500 to 900kPa
Flow rate: 300 L/min or more
Dew point: -20 deg C or less (depend on the specifications)