OX100 Current Limit Type Oxygen Analyzer
Model OX100
Current Limit Type Oxygen Analyzer


Current Limit Type Oxygen Analyzer (General purpose)  Model OX100

The OX100 Oxygen Analyzer measures oxygen concentration from ppm to percent levels and is ideal for ppm-level measurements in such applications as nitrogen reflow furnaces and semiconductor plants.

The zirconia oxygen sensor is based on the limiting current principle and is compact, yet provides high-performance. The sensor is designed for installation by direct insertion.

Where a sample gas contains organic solvent at a few ppm to percent level in applications such as nitrogen reflow furnaces, an optional sampling unit is required.


   Bulletin  for OX100

   User's Manual  for OX100

   Selection Guide  for Oxygen Analyzer
* A single instrument capable of measuring 1000ppm to 25% oxygen
* Simple one touch calibration
* Self diagnostics assure high reliability
* One touch sensor replacement
* Compact size
Measured gas Measures O2 in N2 or mixed gases; can’t contain silicone vapor, flammable gas or organic solvent.
Measuring principle Zirconia current limiting method
Measuring range 0 to 1000 ppm O2, 0.1 to 25% O2 (measuring range autoswitching)
Note: For measurement in the range of 1500 to 4500ppm, use the OX102.
Sampling method Natural diffusion (direct input)
Measurement gas conditions Atmospheric pressure, temperature 0 to 70°C, gas flow up to 1 m/s, no corrosive gases like SO2 (can’t be used in atmosphere with flammable gases, corrosive gases, or silicone vapor.
Output signal 4-20 mA DC isolated
Contact output 2 outputs
Calibration Atmospheric pressure calibration, 1000ppm standard gas calibration
Installation Indoor, panel mounting
Power supply 100 V AC ±10%, 50/60 Hz.
Power consumption Up to 15 VA
Weight Converter; approx. 600 g, sensor; approx 150 g
Linearity

±5% of span (for 0-1000 ppm range, with O2 being in the range 0 to 100 ppm).
±3% of span (for 0-1000 ppm range, with O2 being in the range 100 to 1000 ppm).
±2% of span (for 0.1 to 25% range, with O2 being in the range 100 to 1000 ppm).

Repeatability ±5% of span (for 0-1000 ppm range, with O2 being in the range 0 to 100 ppm).
±2% of span (for 0-1000 ppm range, with O2 being in the range 100 to 1000ppm).
±2% of span (for 0.1 to 25% range, with O2 being in the range 100 to 1000 ppm).
Stability (Drift) ±2% of span per week
Current flowing in the Zirconia fixed electrolyte causes oxygen to move from the negative to positive electrode (acts as an oxygen pump). If the flow of oxygen to the negative electrode is restricted by the alumina substrate then a current saturation area is created and the same current flows regardless of the applied voltage. The current that flows is called the limiting current. When a fixed voltage is applied, the current flow is proportional to the oxygen concentration.

1. Development lab, globe box
2. Clean room for testing, environment, fermentation, biochemicals
3. Continuous monitoring of gas flow in food packing
4. Aging machine