Intelligent Industrial Recorders
µR10000 & µR20000
Features
Calibration Correction (Option, /CC1)
This function enables minute adjustments when a certain correction is necessary due to deterioration of a sensor, or if measured values fall within the accuracy range but fluctuate relative to the displayed values on other instruments. Input signals are corrected based on a specified broken line, and its values are measured.
Number of Set Points
You can select the number of points that make up the segments (including the start and end points) in the range of 2 to 16.
Using revision values
Using absolute values
Corrected value (B4) of the fourth set point = a ? b
Example of the Input Values
| Input Value | Corrected Value | Revision Value | Absolute Value |
|---|---|---|---|
| 9.8° | 10.0° | 0.2° | 10.0° |
| 90.5° | 90.0° | -0.5° | 90.0° |
24 VDC/AC Power Supply (Option, /P1)
The unit can be powered under 24 VDA or 24 VAC.
- Rated power supply: 24 VDC/AC
- Allowable power supply voltage range: 21.6 to 26.4 VDC/AC
- Dielectric strength:
- Power supply to ground terminal: 1000 VAC
- Power consumption (approx.):
| Supply Voltage | Max. |
|---|---|
| 24 VDC | Approx. 50 VA |
| 24 VAC (50/60 Hz) | Approx. 60 VA |
Portable Type (Option, /H5[ ] )
This model comes with a convenient carrying handle and portable power cord.
Names of Parts
![]() µR10000 |
![]() µR20000 |
- Carrying handle
- Protective ground terminal (spare)
- Power terminal with protective ground terminal (attache the specified power cord)
- Feet (four)