Yokogawa's pH and ORP analyzers are highly reliable. They feature advanced functions and are useful for a wide variety of applications including water quality management in a broad range of production processes and in medium-sized wastewater treatment plants, or as general pH and ORP control systems.
Based on Yokogawa's track record and years of experience, a comprehensive range of products has been produced to provide solutions best suited to individual applications.
In addition to the PH8EFP, PH8ERP, OR8EFG and OR8ERG standard Ryton pH and ORP sensors, specialty pH and ORP sensors are available for various applications.
Although pH / ORP sensor was used as an analog detector in the past, it became possible to use it as a digital detector (with Variopin connector) by installing the SENCOM Smart Adapter SA11. pH / ORP sensor is a pH detector that compatible to on-site use and various combinations.
Ryton pH/ORP sensors PH8EFP, PH8ERP, OR8EFG, OR8ERG
1. pH Sensors
1-1. KCl Refillable Type Electrode PH8ERP
Measured object | Hydrogen ion concentration (pH) in aqueous solution |
Measurement principle | Glass electrode method |
Measuring range | pH 2 to 12 |
Measuring temperature | -5 to 80℃ |
Measuring pressure | Atmospheric pressure to 50 kPa |
Temperature compensation sensor | Pt1000 |
Weight | Approx. 0.4kg |
Digital communication function | Support SENCOM4.0 (Variopin connector and SA11 adapter connected) |
1-2. KCl Filling Type Electrode PH8EFP
Measured object | Hydrogen ion concentration (pH) in aqueous solution |
Measurement principle | Glass electrode method |
Measuring range | pH 0 to 14 |
Measuring temperature | -5 to 105℃ |
Measuring pressure | Atmospheric pressure to 10 kPa (General purpose or big volume tank 500 ml) Atmospheric pressure to 500 kPa (Medium pressure) |
Temperature compensation sensor | Pt1000 |
Weight | Approx. 0.4 kg |
Digital communication function | Support SENCOM4.0 (Variopin connector and SA11 adapter connected) |
2. ORP Sensors
2-1. KCl Refillable Type Electrode OR8ERG
Measured object | Oxidation-Reduction potential in aqueous solution |
Measurement principle | Metal electrode method |
Measuring range | -1500 to 1500 mV |
Measuring temperature | -5 to 80℃ |
Measuring pressure | Atmospheric pressure to 50 kPa |
Weight | Approx. 0.4 kg |
2-2. KCl Filling Type Electrode OR8EFG
Measured object | Oxidation-Reduction potential in aqueous solution |
Measurement principle | Metal electrode method |
Measuring range | -1500 to 1500 mV |
Measuring temperature | -5 to 105℃ |
Measuring pressure | Atmospheric pressure to 10 kPa (General purpose or big volume tank 500 ml) Atmospheric pressure to 500 kPa (Medium pressure) |
Weight | Approx. 0.4kg |
Cyanide-bearing wastewater from mining and electroplating facilities and certain types of chemical plants is toxic and must be treated by oxidation with chlorine or chloride to bring the cyanide concentration within regulatory limits.
Industry:Electrical and Electronics
In flue gas desulfurization systems that use magnesium hydroxide (Mg(OH)2) slurry, the consumption of the desulfurization agent (Mg(OH)2) is controlled by using online pH analyzers. A great concern in the pH measurement is heavy staining of the pH electrodes by the Mg(OH)2 slurry. To ensure accurate measurement, frequent cleaning of the electrodes with an acid is required, adding to both maintenance workload and cost.
Industry:Chemical, Power
The lifetime of a pH sensor has a significant impact on the overall annual costs of a pH measuring loop. Optimizing four key factors will decrease these costs and optimize process control and overall plant efficiency.