Introduction
There are a number of apoptosis analysis methods that use imaging. When focusing on nuclei, such phenomena as chromatin aggregation, nuclei condensation and DNA cutting,etc. are observed along with the progress of apoptosis, which are recognizable from microscopic images. The following is an example of the analysis of nuclear transformation by using the images acquired by CV7000 and 「Nuclear Morphology」function of the analysis software.
Analysis Results
Images were acquired using multiple object lenses and analyzed by the analysis software. As a result, changes in the nuclear areas due to apoptosis were observed with any of the objective lenses(Fig. 1). In the created doseresponse curves of three factors (Fig. 2), both the 4X and 40X lenses gave almost the same results. In addition, bar charts of individual cell distribution in the wells with staurosporine concentrations of 0μM and 10μM (Fig. 3) and pie charts (Fig. 4) were created using Spotfire®, both of which clearly indicated apoptosis. In this way, the 「Nuclear Morphology」function of the analysis software enables multilateral analysis of changes in nuclear morphology, and stable analysis of images acquired by as low as 4X magnification object lenses, too.
Fig. 1: Images captured by CV7000, and their recognition results
Based on the nuclear area per cell, live cells and cells with apoptosis are recognized and colored in green and red, respectively.
It is clearly shown that almost all nuclei are fragmented due to apoptosis in the wells with 10μM staurosporine.
Experiments
- HeLa cells were cultured in 96-well plates with 10,000 cells/well for 24 hours.
- Add Staurosporine, fix with formaldehyde four hours later, and stain nuclei with Hoechst33342.
- CV7000 observation conditions are shown below: (wavelength: 405nm)
Magnification Exposure time (405nm) Number of images (per well) Number of recognized cells (well average) 4x 500 msec 1 3,452 cells 10x 250 msec 4 3,028 cells 20x 250 msec 9 1,243 cells 40x 250 msec 16 493 cells -
Analyze captured images using the 「Nuclear Morphology」function of the analysis software in its advanced mode.
• Extract nuclear areas from the images to recognize the nuclei with less than 100μm2 area as micronuclei fragmented due to apoptosis.
• Calculate the area and average fluorescence intensity of the nuclei and fragmented micronuclei, respectively, as well as the number of micronuclei.
Fig. 2: Staurosporine dose-response curves: Imaged with (a)4x or (b)40x lens
Fig.3 Cell number distribution graphs (40X)
Fig. 4 (a): Pie-chart : Ratio of cells at various staurosporine concentrations. (Magnification: 40x)
The ratio of cells with nuclear fluorescence intensity of 800 or more are colored in green.
Fig. 4 (b): Pie-chart: Ratios of cells at various staurosporine concentrations. (Magnification: 40x)
The ration of cells with one or more fragmented nuclear granules are colored in green.
Our Social Medias
We post our information to the following SNSs. Please follow us.
Follow us | Share our application | |
@Yokogawa_LS | Share on Twitter | |
Yokogawa Life Science | Share on Facebook | |
Yokogawa Life Science | Share on LinkedIn |
Yokogawa's Official Social Media Account List
相关产品&解决方案
-
CellPathfinder
CellPathfinder is designed for our HCA systems, CQ1 and the CellVoyager series. From beginners to experts, the analysis software lets you quantify subtle physiological changes and even label-free samples with various graph options.
-
CellVoyager高内涵分析系统CQ3000
CQ3000可以根据应用组合选项,在培养细胞的同时快速获取高分辨率的3D图像。
-
CQ1台式高内涵分析系统
CellVoyager CQ1采用节省空间的台式设计,提供高质量的共聚焦图像和延长的活细胞成像。
-
CV8000高内涵筛选系统
CellVoyager CV8000是先进的高内涵筛选系统。改进的内置培养箱可以让客户分析长期的活细胞反应。凭借其可扩展性、4个摄像头、5个激光器和可选的内置移液器,该系统允许日益复杂的分析开发和高内涵筛选。
-
生命科学
解锁生命本质
横河电机灵敏、高精度的解决方案可实现快速准确的测量,支持突破性的研究,加快药物发现,并优化大规模生物生产。
-
高内涵分析
我们的高内涵分析系统利用强大的分析软件助力从基础科学到药物发现筛选的一系列研究应用。