DO70G荧光法溶解氧传感器

DO70G荧光法溶解氧传感器通常用于水处理工厂,如污水处理及活性污泥废液处理等工艺。

该荧光法传感器响应快速,具有良好的长期稳定性,传感器帽和电缆可更换,便于维护。传感器中还包含可进行温度补偿的温度传感器。DO70G传感器可用于PB350G/PB360G浮球式支架或DOX8HS浸入式支架。

详细介绍

  • 荧光法DO传感器使用相位差测量法,可在较长时间内实现稳定测量。
  • 测量范围:025 mg/l
  • 传感器帽可由任何人轻松更换。
溶解氧 0 to 25 mg/l (ppm)
液体温度 0 to 50℃
液体压力 0 to 500 kPa
流速 20 cm/s or more
温度补偿传感器  22k NTC
电缆长度 3, 5, 10 m
重量 传感器;约0.1 Kg
电缆;3 m:约0.4 Kg5 m:约0.6 Kg10 m:约1.1 Kg  
重复性 0.1 mg/l或满量程的3%,取较大值。
温度补偿误差  040℃的范围内温度变化±5℃,误差不超过满量程的±3% 
响应时间 不超过120(90%响应)

参考

概述:

In a batch tank, analyzers such as dissolved oxygen (DO) analyzers, oxidation-reduction potential (ORP) meters, and mixed liquor suspended solid (MLSS) analyzers are used. To reduce the maintenance frequency, measures must be taken to prevent dirt accumulation on the sensor. As the liquid level in a tank can vary greatly, great care should be taken when selecting the sensor installation position in a tank.

概述:

When used to measure the dissolved oxygen in an aeration tank, sensors tend to quickly become contaminated and need frequent cleaning.

概述:

Removal of free oil and grease from a wastewater stream reduces the potential for equipment problems to occur further downstream. There are three forms of oil encountered in wastewater treatment at a refinery. 