What are the burnout determination criteria (just for thermocouples)?

4.2 µA flows from a burnout power supply, resistance is measured, and the determination is made.

: Thermocouple is normal: Less than 10 KΩ
Thermocouple breaks, and while it is breaking: 10 KΩ or more

相关产品&解决方案

  • DR130

    The DR130 can record data in 10 colors from all measurement points at 2-second intervals. It has a variety of recording functions including analog trend recording over an effective recording width of 150 mm, recording of digital measured values, recording of various messages, zone recording, and partially compressed and expanded recording, thus enabling data to be read off speedily.