There's no output with an RTD temperature converter. What should I do?

  1. Check the power supply wiring.
  2. Check whether the power supply voltage meets the specifications.
  3. For Pt100 input specifications, connect a 100 Ω resistance between input terminals A and B, short the B terminals, then confirm that the output is equivalent to 0°C.
  4. If the output signals is of the current specifications, attach an external 250 Ω resistance and measure the voltage.

Productos y Soluciones Relacionadas

  • JUXTA VJ Series Limit Alarms

    The VJ series limit alarms are compact, plug-in type alarm setting devices with easy-to-use functions. Four models are available: the VJAK for 2-wire system transmitters; the VJHK with DC voltage/DC current inputs; the VJRK with RTD inputs; the VJTK with thermocouple inputs; and the VJUK with universal inputs.

    Leer Más
  • JUXTA VJ Series Signal Conditioners

    Compact size, plug-in style and microprocessor based high performance capabilities (Multi-function types) makes the VJ Series Signal Conditioners an excellent choice for almost every application. Analog output, Hi-Low alarm relay outputs or RS-485 MODBUS communication is available as optional second output.