Fast, Simple, and Low-Cost Hydrogen Measurement

General H2O electrolysis process
Measuring hydrogen concentration in a typical H2O electrolysis process has the following challenges:
(1) Conventional thermal conductivity meters and alloy thin film meters; Slow response, tens of seconds to 60 seconds
(2) The anode side need to multiple pretreatments for water separation and dehumidification.
In H2O electrolysis plants that produce hydrogen, cross-leakage to the anode and cathode sides due to membrane degradation is an issue for safety management, and the inability to achieve a fast response in general hydrogen concentration measurement is a major issue for safety monitoring.
Since the gas components of the H2O electrolysis process are hydrogen and oxygen, it is possible to estimate the gas concentration by gas density. The applicability of gas density meters in hydrogen concentration measurement is introduced in this document.

Fig. 3 Schematic diagram of hydrogen measurement
Advantages of Using a Gas Density Meter for Hydrogen Measurement
Not only the analyzer's fast response but also pre-processing can be simplified, achieving overall fast response
- Simplified Plant Design
simple pre-processing contributes to the simplification plant design - Fast Response
The gas density meter is faster than conventional ones
Product Details

GD80
Gas Density Meter Detector

GD802G
Gas Density Meter Converter
Principle
Two pairs of piezoelectric elements for drive and detection are attached on four points of a thin-wall cylindrical resonator. A platinum temperature sensor is mounted to the density sensor.
The sample gas from inlet passes over the outside of the thin-walled cylindrical resonator and is exhausted at the gas outlet.

Fig. 4 Detector structure
Gas Density Meter Specifications
| Sample Gas Conditions | -Sample Gas: All gases except for corrosive gases (Contact Yokogawa for details.) -Temperature: -10 to 60℃ (No condensation) -Flow Rate: 0.1 to 1 L/min -Pressure: 50 kPa to 600 kPa abs. |
|
|---|---|---|
| Principle | Multi-mode oscillation technology | |
| Performance Specification | -Range: Concentration: 0 to 100 vol% -Minimum Range: Equivalent to 0.100 kg/m³ density -90% Response Time: Approx. 5 s (Time required for the reading to reach 90% of its final value after the indicated value begins to change (at a gas flow rate of 0.6 L/min)). -Linearity: ± 1% -Repeatability: ±0.5% or concentration equivalent to ±0.001 kg/m³ density (Whichever is greater. ) -Zero / Span Drift: Equivalent to ±0.003 kg/m³ density/month (Provided that the detector, converter, and pressure transmitter are not affected by ambient temperature.) |
|
Related Industries
-
수소
청정 수소 시대가 열렸습니다.
Related Products & Solutions
-
Gas Density Meter GD80, GD802G, GD802S
Fast-response gas density measurement for safe and stable operation. More than density, monitor gas concentration, molecular weight, specific gravity, hydrogen purity, and replacement concentration.
-
Gas Density/Hydrogen Purity Analyzer
가스 밀도 검출기 및 분석기는 가스 밀도의 연속 측정을 제공합니다. 뿐만 아니라 비중 및 분자량을 포함한 여러 가지 중요한 파라미터를 제공합니다.
-
In-Situ Gas Analyzer TDLS8000
Yokogawa의 새로 TDLS™8000은 업계 최고의 기능을 하나의 견고한 장치에 모두 탑재하고 있습니다. 이 플랫폼 설계는 샘플 추출 및 컨디셔닝이 필요 없는 현장 측정을 위한 것입니다.
-
프로브형 가변 다이오드 레이저 분광기 TDLS8200
별도의 Sampling 장치가 불필요하고 installation flange가 한쪽에만 있으며, 고속 측정/고속 응답을 유지함과 동시에 설치 비용이 반으로 절감되어 기존 장비를 손쉽게 교체할 수 있습니다.