Measurement and Monitoring of O2 Concentrations in Steel Mill Reheating Furnaces

다운로드 (622 KB)

Introduction

A furnace for heating slabs needs to be operated under low oxygen conditions at high temperatures of 1000℃ and above to prevent oxidation of the steel. The measurement of oxygen concentrations in the furnace is essential in this process.

The ZR22G/ZR802G Zirconia Oxygen Analyzer requires no sampling system and utilizes a long-life sensor, and is therefore ideally suited for measurement of oxygen concentrations in a hightemperature environment. The ZR22G detector with the ZO21P high temperature probe adapter can be used in gas furnaces at temperatures up to 1400℃.

Expected Benefits

  • Maintains high product quality by monitoring oxygen concentrations in steel mill reheating furnaces
  • Ensures stable, continuous measurement of oxygen concentrations
  • Reduces operating costs

Process Overview

Slabs are loaded on a conveyor and fed to a reheating furnace, which is heated to between 900 and 1400℃ by burners that use fuels such as by-product gas. In order to keep the oxygen concentration between 1 and 2%, it is monitored in the preheat, heat , and soak zones. The slabs are heated uniformly without being oxidized, and then sent to the next processing stage.

Reheating Furnace Process

Solution Details

Field Data

Process conditions
Measurement point: furnace reheat, heat, and soak zones
Temperature: 900°C to 1400°C
Pressure: 30 to 50 kPa
Dust: Trace
Fuel: Gas

Measurement system

Auxiliary Ejector for High Temperature Use

Detector: ZR22G-015-S-Q-☐-☐-☐-E-A/CV
High temperature probe adapter: ZO21P-H-☐-☐☐☐-☐*B
Auxiliary ejector for high temperature use: E7046E☐
Converter: ZR802G-☐-☐-N-N/☐
Flow setting unit: ZA8F-☐*C
Calibration gas unit
Pressure regulator for gas cylinder: G7013XF (inlet W22, outlet Rc1/4) or G7014XF (inlet W22, outlet 1/4NPT)
Case assembly for calibration gas cylinder: E7044KF

Note: the calibration gas cylinder must be purchased locally.
 

Probe Adapter

Utilities
Power supply: 85 to 264 V AC, 50/60 Hz
Instrument air (reference gas): pressure, 300 to 700 kPa

Notes

The material selected for the high temperature probe adapter depends on the temperature of the sample gas. Follow instructions carefully when installing the probe.

SUS: Sample gas temperature ≤ 750°C
Probe orientation: vertical or horizontal
Inconel: 750°C < sample gas temperature ≤ 900°C
Probe orientation: vertical or horizontal
SiC: 900°C < sample gas temperature ≤ 1400°C
Probe orientation: vertical

Note:
When the probe is installed horizontally, bracings must be provided for mechanical support

Probe-Bracings

 

업종

  • 철강

    철강 산업에서는 제품뿐만 아니라 제조 및 운영 기술의 품질을 향상시키고 환경 및 에너지 효율 문제를 해결하는 것 또한 중요합니다. Yokogawa는 고객이 이상적인 플랜트를 만들 수 있도록 돕고 동반 성장을 위해 발전해 나가고 있습니다.

    See More

Related Products & Solutions

  • 지르코니아식 산소/습도 분석기 ZR22G, ZR802G

    이 장비는 고온에서 산소 및 습도 측정이 가능하며 매우 정교한 지르코니아 센서가 장착된 ZR22G 검출기와 디지털 통신 및 데이터 로그 기능을 갖춘 변환기 ZR802G로 구성됩니다. ZR22G 및 ZR802G는 샘플링 장치 없이 연속적으로 측정할 수 있습니다. 센서의 성능을 자가진단하고 (교정 가스를 사용하지 않음) 현장에서 Heater assembly를 교체하여 유지 보수를 용이하게 하여 작업 비용을 줄일 수 있습니다.

     

    See More
  • Oxygen Analyzers

    산소 분석계는 연소 제어, 공정 품질, 안전 및 환경 응용 분야에서 중요한 측정을 제공합니다.

    See More

Top