What are the maintenance parts for the RC800D (Reagent Type Residual Chlorine Sensor Unit)?

Parts No. Name Remarks QTY Recommended Replacement Interval
- Primary acetic acid (CH3COOH)
grade: over 99.0+%
For reagent preparation -  
- Sodium acetate (CH3COONa)
grade: over 98.0%
For reagent preparation -  
- Potassium bromide (KBr)
 grade over 99.0+%
For free chlorine measurement reagent (-F, -C) -  
- Potassium iodide (KI)
grade over 99.0+%
For total chlorine measurement reagent (- T) -  
K8005JC Measurement Electrode (rotating electrode) gold-alloy electrode 1 yearly
K8005UH Reference electrode Platinum electrode (-F, -T) 1 (*1)
K8005UC Reference electrode Silver/Sliver chloride electrode for combined chlorine insensitive type (-C) 1 (*1)
K9332ZJ Glass beads 2 Packs/Piece 1 yearly
K9088PE Polishing Powder (Alumina) For polishing measurement electrode 1 -
K9332SR Rotating contact Part for the electrode unit 1 yearly
K8005LB Driven shaft assembly Part for the electrode unit 1 3 years
Y9115XB O-ring Part for the electrode unit 1 3 years
K8005LC Motor Part for the electrode unit 1 3 years
B1005AC Coupling Part for the electrode unit 1 3 years
A1633EF Fuse 250 V/2.5 A for the smart unit 1 -
A1624EF Fuse 250 V/3.15 A for the terminal box (/ TB option) 1 -
K9720FZ Sand for filter 1 L for the sand Filter (-2N,-3N,-2A,-3A) 1 yearly
K9087XA Air Pump (100 V) For the Air pump (- P1) 1 3 years
K9087XF Air Pump (110 V) For the Air pump (- P2) 1 3 years
K9087XG Air Pump (200 V) For the Air pump (- P3) 1 3 years
K9087XH Air Pump (220 V) For the Air pump (- P4) 1 3 years
K9041RA Lubricating oil For liquid pump monthly maintenance 1 1 month (*2)
L9819AA Bellofram For reagent flow control of liquid pump 1 6 Months
L9819AB Bellofram For sample flow control of liquid pump 1 6 Months
K9041HC Valve seat For liquid pump 1 -
K9041SU Valve seat For liquid pump (4 pcs of K9041HC) 1 6 Months
L9862AY Activated Charcoal Filter For zero calibration 1 yearly
K9332NN Filter Parts of the sand filter (-2N,-3N,-2A,-3A) 1 yearly

*1:Replace it when it is damaged.
*2:Oil is in a bottle, so needs to be applied every month, but not necessary to buy every month a bottle. It will last 3 years.

Productos y Soluciones Relacionadas

  • Liquid Analyzer for Turbidity and Chlorine FLXA402T

    The FLXA402T is the next-generation analyzer for the Turbidity detector and Chlorine sensor. The FLXA402T can connect up to two sensors of Turbidity, Chlorine, pH, and Conductivity, with digital communications availability.

  • Reagent Type Residual Chlorine Sensor Unit RC800D

    The RC800D is a next-generation digital Turbidity analyzer and connect to FLXA402T. The RC800D can measure free chlorine or residual chlorine (total chlorine). It is suitable for free chlorine measurement under challenging and residual chlorine measurement.

  • Turbidity/Chlorine SMART Sensor Platform

    SENCOM 4.0 technology platform gives you more insight with reliable measurement by digital communication. This has been extended to Turbidity and Chorine measurement technologies.


Top