I’m inputting a signal of 0–150°C/1–5 V from a transformer to the µR10000. The temperature does not exceed 100°C, so can I scale it to the chart paper values of 0–100 (so that the chart records 100 when the temperature is 100°C)?

Yes. Set the range settings as follows.

Mode: Linear scaling
Range: 6 V
Left span: 1.000 V
Right span: 3.667 V
Left scale: 0.0
Right scale: 100.0

(Note: if the mode is 1–5 V, you cannot set the right span outside of 4.8–5.2 V.)

 

Related Products & Solutions

  • µR10000 & µR20000

    The μR series are the compact industrial recorders with the recording widths of 100 mm (μR10000) and 180 mm (μR20000). Measured data can be saved to an SD memory card while recording on chart paper.

    See More

Top