WZ-H Cable Signal Transmitter
WZ-H cables are compatible with all ZR22, ZR402, and AV550G model Oxygen Analyzer Systems. The purpose of these cables is to transmit the signal from the sensor/detector to the analyzer. Additionally, the cabling provides for electrical input to the heater assembly within the detector for cell temperature control.
WZ-H-6S & 6H High-Temperature Cable
Nominal Insulation Thickness FEP- 0.007 inches 0.178 mm
Nominal Jacket Thickness- 0.012 inches 0.81 mm
Nominal Outside Diameter- 0.295 inches. 8.26 mm
Number of Conductors- 7
Number of Conductor Pairs- 3 + single drain
AWG size- 16
Conductor Strand- 19/.0117
Conductor: 3 Pair Shielded CL2P CMP AWG 16/19 x 0.0117” TC FEP/ FEP
Insulation: FEP 0.007” Wall min. O.D. = 0.071” to 0.073”
Shield: Aluminum / Mylar Shield with AWG18/7 X 26 TC Drain Wire
Jacket: Gray FEP 0.012” wall Temperature Range: -80°C to 200°C (-122°F to 392°F)
Braid: None
Voltage: UL 300V Power Limited
Capacitance: 22pF/ft Nominal Weight: 67lbs. /1000ft.
Compliances: NEC Article 725-CL2P NEC Article 800-CMP PCC FT6-C(UL) (CMP only)
WZ-H-3H High-Temperature Cable
Nominal Insulation Thickness FEP- 0.006 inches 0.152 mm
Nominal Jacket Thickness- 0.010 inches 0.254 mm
Nominal Outside Diameter- 0.188 inches 4.775 mm
Number of Conductors- 3
Number of Conductor Pairs- 1 + single drain
AWG size- 14
Conductor Strand- 19/.0147
Conductor: 2 Conductor, Shielded CL2P CMP AWG 14/19 x 0.0147” Tinned Copper FEP/FEP
Insulation: FEP 0.006” Wall min. O.D. = 0.082” to 0.084”
Shield: Aluminum / Mylar Shield with AWG16/19 Tinned Copper Drain Wire
Jacket: Gray FEP 0.010” wall, O.D. 0.188” (0.225” max)
Temperature Range: -80°C to 200°C (-122°F to 392°F)
Braid: None
Voltage: UL 150V Power Limited
Capacitance: 42pF/ft Nominal
Weight: 42lbs. /1000ft.
Compliances: NEC Article 725-CL2P NEC Article 800-CMP C(UL) (CMP only)