What factors should be considered on the selection of the conductivity sensor?

  1. Sensor material: 
    Sensor materials are very significant; the insulator between the two electrodes must not deteriorate over time. For example, when measuring in Ultra-Pure Water (UPW) is being made at 18MΩ: any of the insulation degradation would result in higher than actual conductivity measurement.
     
  2. Compatibility with the process:
    All the materials must be compatible with the process. They must not rust, swell, shrink, etc.
     
  3.  Robustness:
    The sensor must be as strong and robust as possible. The geometric dimensions between the two electrodes are critical: any change will significantly impact the conductivity reading.
     
  4. Temperature sensor accuracy:
    The temperature sensor must be of high quality; a one °C error could result in a 10% conductivity error.

 

For detailed information please check our Technical Information: Contacting Conductivity sensor selection guide. 

Productos y Soluciones Relacionadas

  • Conductivity Sensor SC4AJ

    The SC4AJ conductivity sensor has a convenient, compact design and is widely used in various applications including boiler water and hydroponics conductivity measurement and pure water resistivity measurement in the semiconductor, food, pharmaceutical, and power industries.

  • Contacting Conductivity Sensors SC42/SC4A(J)

    The measurement of specific conductivity in aqueous solutions is becoming increasingly important for the determination of impurities in water.  Yokogawa has designed a full range of precision sensors and instruments to cope with these measurements, even under extreme conditions.

  • High Conductivity Sensors ISC40/SC42

    The measurement of specific conductivity in aqueous solutions is becoming increasingly important for the determination of impurities in water. Yokogawa has designed a full range of precision sensors and instruments to cope with these measurements, even under extreme conditions.


Top