Conductivity Sensor Cleaning

Soak the sensor in 5 - 10% Hydrochloric acid (HCl) in water solution. for 5 - 10 minutes, then rinse. If you have trouble finding 5 - 10% HCL you can buy muriatic acid at a building supply house and it is usually 20 - 30% HCL. Check the label. Dilute it with tap water to get something close to 10%. 

BE VERY CAREFUL WHEN DILUTING THE ACID. USE PROTECTIVE CLOTHING (GLOVES, FACE SHIELD ETC.) ALWAYS ADD ACID TO WATER, NEVER ADD WATER TO ACID.

Related Products & Solutions

  • Conductivity Sensors

    • Conductivity sensors and electrodes
    • Process conductivity, resistivity, WIFI, demineralizer water, RO water, percent concentration, boiler blowdown, and TDS 
    • Retractable, flow thru, immersion, and direct insertion
    See More