SC200/202/400/402/450/FLXA21/FLXA202/DC400/DC402 Testing or simulating a contacting conductivity input

To test the conductivity input do the following: 

  1. Write down the cell constant and change it to 1.00 (this is not mandatory, it makes the math easier) 
  2. Place a jumper wire between 13 to 14 and another one on 15 to 16 (not necessary on the DC400/402) 
  3. Place a resistance between the 13/14 and 15/16 connection. If the unit is set in resistivity it should match the resistance. If it is displaying conductivity it should read Y where Y = (cell constant)/(input resistance) and the units are Siemens. So if the result is 0.005 Siemens this equals 5 milliSiemens.and 0.000005 Siemens equals 5 microSiemens. 

Remember to set the cell constant back to the original value if you changed it.

Related Products & Solutions

2-Wire Transmitter/Analyzer FLXA202/21

The FLEXA™ series analyzers are used for continuous on-line measurements in industrial installations. With an option for single or dual sensor measurement, they are the most flexible two-wire analyzer available.

DC400

Model Name: DC400

SC200G

Model Name: SC200G

SC200S

Model Name: SC200S

SC202G

Model Name: SC202G

SC202S

Model Name: SC202S

SC202SJ

Model Name: SC202SJ

SC400G

Model Name: SC400G

SC402G

Model Name: SC402G

Conductivity Analyzers

Conductivity meters, analyzers and transmitters are used for continuous process measurement and monitoring of conductivity, resistivity, WIFI, demineralizer water, RO water, percent concentration, boiler blowdown and TDS.

Liquid Analyzers

Liquid analyzers are used for monitoring process chemistry including water quality, providing process optimization and control.

Process Analyzers

Process analyzers are used to determine the chemical composition or physical properties of substances involved in industrial processes.

Top