Extractive Tunable Diode Laser Spectrometer TDLS8220

Built upon Yokogawa's second generation of TDLS analyzers, the TDLS8220 continues to house all the industry's leading features in one robust device. The improved platform is designed for accurate, low maintenance, and reliable extractive oxygen measurements of process gas streams. 

Features

  • Analyzer, flow cell, heat trace control, insulation, purge control, and wiring termination pre-integrated
  • Drop-in replacement for existing analyzer technologies
  • Integrated tube-in-shell flow cell allows for tight sample heating control
  • Pressure and temperature compensation included
  • Advanced diagnostics with 50 days of data and spectra storage

Reliability

Auto gain graphs

Reference Cell
Internal reference cell in the laser module ensures peak locking during trace measurement.

 

Auto gain
Auto-gain enables wide signal ranges against dynamic variation of transmission.

 

Validation
Validation can be initiated manually, remotely, or automatically on a daily, weekly or monthly basis defined by the user.

TruePeak

With TruePeak we can measure the area of the absorbance peak. This eliminates effects from changing background gases, allowing for simple pressure and temperature compensation.

TruePeak spectra wavelengthsTraditional TDL spectra wavelengths ​​​​​

Intuitive Touchscreen HMI

  • Touch screen 7.5 inch color LCD on HMI
  • Simple operation
  • Provides full analzyer control and access to monitoring, historical files, and confiugrations
  • Remote installation with the ability to connect up to 4 different TDLS analyzers
 

HMI Trend Display

Details

Standard Specifications

Measurement object O2
Measurement system Tunable diode laser spectroscopy
Measured components and range Measured component Min. range Max. range
O2 0-1% 0-25%

Output signal
(Analog Output)

2 points, 4 to 20 mA DC
Output types; Gas concentration, Transmission, Process gas temperature, Process gas pressure
Output range; 3.0 to 21.6 mA DC
Digital communication HART, Ethernet, Modbus/TCP

Contact outputs
(Digital Output)

2 points, contact rating 24 V DC, 1 A
DO; Function: Activate during Warning / Calibration / Validation / Warm up / Maintenance conditions
Fault; Function: Activate during normal condition, not activate during fault condition or when the system power is off
Valve control output 2 points
Function; Activate calibration or validation solenoid valves for zero, span or validation gas
Output signal; 24 V DC, 500 mA Max. per terminal
Alarm Warning;
Gas concentration low/high, Transmission low, Process pressure low/high, Process temperature low/high, Validation required, Validation failure, Zero/Span calibration error, Non process alarm, External alarm
Fault;
Laser module temperature low/high, Laser temperature low/high, Detector signal high, Peak center out of range, Reference peak height low, Absorption too high, Transmission lost, Reference transmission low, Reference peak height high, Laser unit failure, Laser module error, File access error, E2PROM access error.

Contact inputs
(Digital input)

2 points
Function;
External alarm/Calibration start/Validation start/Stream switch
Contact specification;
Zero voltage contact input
Input signal; Open signal;
100 kΩ or more, Close signal; 200 Ω or less

Input signal
(Analog input)

2 points, 4 to 24 mA DC
Input types; Process gas pressure, Process gas temperature
Self-diagnostics Laser Unit temperature, Sensor Control Unit temperature, Laser temperature, Detector signal level, Memory read/write function, Peak locking condition
Calibration Calibration method; Zero/Span calibration
Calibration mode; Manual, Auto, Local initiate (HMI)
Validation Validation method; Up to 2 points
Validation mode; Manual, Auto, Local initiate (HMI)
Warm-up time 5 min.
Protection degree IP65
Hazardous area classifications The analyzer is designed fo operation in hazardous area for gaseous release. Class 1 Division 2 Groups A, B, C, D with purge system. 
Sample gas condition Sample gas temperature: Max 120°C
Sample gas pressure: 90 to 500 kPa absolute
Installation condition Ambient operating temperature: -20 to +55°C
Humidity: 0 to 95%RH at 40°C (non-condensing)


YH8000 HMI Unit

Display Touchscreen 7.5 inch TFT color LCD panel, 640 x 480 (VGA)
Communication Ethernet; RJ-45 connector, Communication speed; 100 Mbps
Protection degree
of enclosure
IP65, NEMA Type 4X
Weight 4 kg
Mounting Analyzer mount (Front, left-side, right-side) with tilt function, Pipe mount or Panel mount
Cable Entries 1/2NPT or M20 x 2
Installation
conditions
Ambient operating temperature; -20 to 55℃
Storage temperature: -30 to 70℃
Humidity: 10 to 90%RH at 40℃ (Noncondensing)
Power Supply 24 V DC ±10%
Hazardous area
classifications
Division 2, Zone2: Non-Incendive/Type n; FM, cFM, ATEX, IECEx

Resources

Overview:

Enhance the efficiency and safety of your vapor recovery systems and flare/vent headers with Yokogawa's cutting-edge Tunable Diode Laser Spectrometers – the TDLS8000 and TDLS8220. Elevate your oxygen measurement capabilities with our interference-free analyzers, providing direct measurements. Experience unparalleled reliability, ensuring minimal upkeep and secure operations without the need for system shutdowns.

Overview:

Enhance the efficiency and safety of your vapor recovery systems and flare/vent headers with Yokogawa's cutting-edge Tunable Diode Laser Spectrometers – the TDLS8000 and TDLS8220. Elevate your oxygen measurement capabilities with our interference-free analyzers, providing direct measurements. Experience unparalleled reliability, ensuring minimal upkeep and secure operations without the need for system shutdowns.

Downloads

Looking for more information on our people, technology and solutions?

Contact Us

Top