Measuring Process Oxygen with Truepeak TDLS In Lieu of Paramagnetic

Oxygen for the following processes and applications are traditionally measured using paramagnetic technology, but the advent of Truepeak TDLS has proven the paramagnetic measurement to be inferior.

Table 1: Oxygen process applicationsProcess oxygen is the desired process control measurement  in many manufacturing processes, such as: 

  • Ethylene oxide (EO)
  • Formaldehyde
  • VCM/EDC
  • Isopropyl alcohol (IPA)
  • TiO2
  • Isocyanates
  • Acrylonitrile
  • Acetonitrile
  • PTA
  • Refining, etc

Oxygen is also monitored for safety in areas such as:

  • Vent headers
  • Flare headers
  • Tank gas blankets
  • Ship/barge loading areas (marine vapor recovery)
  • Chemical processing areas 
  • Refining areas
  • Hydrogen purity areas (chlorine cells), etc

Measurement Principle and Molecular Interference 

The paramagnetic measurement of oxygen is considered the second principle since it produces oxygen concentration as a function of a measured magnetic effect. 
In the case of the magnetodynamic, or “dumbbell,” analyzer, the oxygen concentration is obtained from the current required to return a magnetically deflected dumbbell to a neutral position. For thermodynamic, or “magnetic wind,” analyzers, sample convection measured by thermistors linked to a Wheatstone bridge gives a differential resistance, which is correlated with oxygen concentration. Differential-pressure type analyzers use the signal generated from a Wheatstone bridge, which measures the degree of cross-flow in reference gas channels connected in parallel. The cross-flow is produced as a result of one reference channel being introduced to the sample gas in the presence of an electromagnet; this creates reference gas flux, which is measured by a thermal flow sensor. 

The result in all cases is a measurement that is susceptible to magnetic interference from other molecules. Lists of molecules are published and often listed in the manuals of various analyzer manufacturers. There are well over 100 known interfering molecules, and the list is steadily growing. Truepeak TDLS is unaffected by interfering molecules because it measures oxygen using first principles. The molecules absorb the laser's light it makes contact with, and the resulting light loss is measured. 

Sampling System Effect on Wet Gas Analysis 

Paramagnetic analyzers require the use of a sampling system to obtain a clean, dry sample. The dry measurement would cause an undefinable false-high reading in-process oxygen if any moisture were present in the original sample. When used in-situ, TDLS will give a wet measurement, which is more representative of process conditions. Extractive Truepeak TDLS is also capable of wet measurement as long as the sample gas is non-condensing. 

Thermal Conductivity Compensation 

Paramagnetic analyzers that utilize a "Wheatstone bridge" rely on compensating for temperature and compensating for the sample gas's thermal conductivity values. This is due to the nature of the "Wheatstone bridge" circuit design in which resistors heat an incoming sample; this leads to the gas losing some of its magnetism according to Curie's law. As a result, cooler, more magnetic oxygen is drawn toward the sensor, creating a "magnetic wind." Thermistors are then used to measure oxygen concentration by detecting the rate of heat loss. The sample gas's heat capacity must be determined by analyzing thermal conductivity to accurately measure the "magnetic wind." Thermal conductivity variations in the sample gas have no effect on the Truepeak TDLS measurement.

Reliability 

There are internal moving parts that must be disassembled for cleaning in a paramagnetic analyzer. The TDLS analyzers have no moving parts, and isolation windows can be removed for cleaning during operation. 
The magnetodynamic paramagnetic analyzer’s measuring cell is constructed of a precisely positioned glass dumbbell filled with inert gas and suspended with a taut platinum wire. This delicate arrangement is easily damaged by corrosive gases, pressure spikes, and liquids. Both extractive and in-situ TDLS measurements have only light contact with the measurement area, and as a result, are unsusceptible to damage.

Sample Filters 

An important note is that response time will suffer proportionally as filters clog in an extractive sample system. In-situ TDLS maintains a constant response time, which is unaffected by the fouling of filters and keeps a typical response time of five seconds. Filter failure can result in measurements that can be not only costly but unsafe as well. 

Flow Rates

Low flow rates must be carefully metered in paramagnetic analyzers to get an accurate measurement and prevent sensor damage in some cases. The Truepeak TDLS measurement is immune to the error caused by flow variations and can accept much higher flow rates, which dramatically reduces response times in extractive systems.

Onboard Diagnostics and Calibration 

Many paramagnetic analyzers offer no onboard diagnostic functions or data storage. Performing a validation check on a paramagnetic analyzer is only an indicator of current performance. Yokogawa’s TDLS offers onboard diagnostics to help catch problems before they occur and stores up to 50 days of historical data and spectra. This allows current, as well as past, analyzer data to be validated. 

Calibration must be performed monthly on most paramagnetic analyzers. The Yokogawa TDLS offers online validation for in-situ systems and automated zero and span calibration checks for extractive systems.

TDLS vs Paramagnetic Process Oxygen Measurement
 

Industries

  • Refining

    In the ever-changing marketplace, refineries are seen not only as crude processing units but also as profit centers. At the same time, there is a keen awareness of the need for safety at such facilities. A total production solution that encompasses planning, scheduling, management, and control is required to achieve long-term goals for profitability, efficiency, and environmental protection. With years of expertise in the automation field, Yokogawa can bring you affordable total solutions for improved operability and a cleaner world.

    See More

Related Products & Solutions

  • Extractive O2 Analyzer TDLS220

    The TDLS220 is specifically designed for accurate, reliable and low maintenance measurement of volume percent (vol%) oxygen (O2) for safety and process applications. It is a viable alternative to paramagnetic O2 analyzers.

    See More
  • In-Situ Gas Analyzer TDLS200

    The TDLS200 is designed to make fast, accurate measurements of gases including oxygen (O2), carbon monoxide (CO), methane (CH4), moisture (H2O), carbon dioxide (CO2), ammonia (NH3), and hydrogen chloride (HCl). It is ideally suited for environments involving changing pressure or temperature.