What is the difference between TDLS Calibration and Validation?

Video FAQ

Questions

  • Why is no field calibration needed?
    • The benefits of TDLS measurements: Drift is negligible and does not affect analyzer performance.
  • Why online validation?
    • Ensure proper analyzer functionality
    • Must for analyzers without reference cells
    • Must for low concentration applications
  • Frequency of validation checks?
    • Depends on the application.
  • Is validation an accuracy check?
    • No, it is a functionality check. 
  • Purpose of validation checks?
    • Analyzer response
    • Peak position check
  • What are the validation steps?
    • Before process snapshot is taken
    • Validation gas introduction and after stabilization a during validation snapshot is taken
    • The non-process area is then purged and an after validation snapshot is taken.
    • The analyzer will display the pass/fail results