Notice: You are currently viewing the website. View the website instead »

Touch Screen GP10/GP20

Industrial processes and test laboratories require rigorous, continuous monitoring in the design, validation, and service stages of product development, product testing, and manufacturing. The GP series is Yokogawa's latest laboratory recorder, with handle and footstands to enable portable and benchtop use.

The GP series is an industry-first multi-point touch panel, to improve intuitive and smart operator control. Users can scroll, pan, zoom historical data, and even write freehand messages on its dust-proof and water-proof display. It delivers industry leading reliability and measurement accuracy.  The GP series is ideally suited for monitoring or troubleshooting long term physical or electrical trends.

Smart User Interface

Variety of display functions

  • Standard display screens clearly indicate channel data with units and tags, plus alarm log, message log, and other event log information.
  • New Custom Display function (option) lets you place graphic images on screens.

Touch screen for intuitive operation

  • The new touch screen provides swipe, pinch in/out, and click and drag controls for fast navigation of common display mode functions.
  • Use "2-point touch" to zoom in or out in the direction of display span.

Supports freehand messages

  • Operators can write a free hand text message directly on the touch screen with a stylus or finger.

Smart Architecture

Add I/O modules when you need more channels

  • The fully modular I/O design allows you to build the system with the channels you need today, and expand the unit's capability any time in the future by purchasing and installing additional modules.

Wide-ranging input/output specifications

  • The 10 channel universal input module measures DC volts, thermocouple, RTD, and contact input signals. A 16 channel digital input module and a relay output module are also available.
    • GX90XA Analog input module: DC voltage, DC current, thermocouple, RTD, contact
    • GX90XA-04-H0 High Speed AI Module:[R4] New!  DC voltage, DC current, thermocouple, RTD, contact
    • GX90XA-06-R1 4-Wire RTD Module:[R4] New!  4-wire RTD, 4-wire resistance
    • GX90YA Analog output module:[R4] New!  Current output (Isolated between channels)
    • GX90XD Digital input module: Remote control input or operation recording
    • GX90YD Digital output module: Alarm output
    • GX90WD Digital input/output module: Remote control input or operation recording/alarm output
    • GX90XP Pulse intput module: Pulse signal data acquisition, integral count
    • GX90UT PID control module:[R4] New!  PID control (2 loop)

Multichannel I/O

  • GP10 and GP20 support up to 30 channels and 100 channels of input respectively.
  • Up to 450 inputs available with an expansion unit that provides easy Ethernet connection to the recorder.

Supports 12 VDC power (GP10 only)

  • Supports 12 V power so you can also use it inside automobiles.

Smart Functionality

PID control  [R4] New!

  • Enables PID and program control (up to 20 loops per system)
  • The web application enables remote operation and monitoring from a browser.
  • Various pre-configured control screens and display are available.
  • Setpoint program control function (up to 99 patterns)

Dual interval measurement  [R4] New!

  • Users have the ability to choose two different scan intervals on a single GP system.

Direct output to printers

  • You can print out reports and snapshots directly from the GP without going through a PC.

Convenient report creation function

  • Reports can be created automatically using a spreadsheet template created in Excel. You can also output to a PDF file.

Browser-based real time monitoring

  • A web browser can be used to fully access the GP using two operating modes— Data and Configuration.

21 CFR Part 11 support (option)

  • Supports the U.S. FDA 21 CFR Part 11 (for pharmaceutical manufacturing).

Easy connection with WT series powermeters (option)

  • Easily acquire and record measured data from power measuring instruments (WT series, power analyzers).

DARWIN compatible functions (Ethernet, RS-422/485, RS-232)

  • Upgrade smoothly from DARWIN data acquisition instruments.

Multi-batch Function (option)

  • Recorder pre-defined channel groups to separate data files with independent start and stop control.

SLMP Communication (Mitsubishi PLC) (option)

  • Protocol function that enables connection from a GP to Mitsubishi Electric PLCs without sequence programs.

OPC-UA Server (option)

  • Data acquired by the GP can be accessed through Ethernet communication from a host system (OPC-UA client).

Aerospace Heat Treatment Supports heat treatment application AMS2750/NADCAP

  • Calibration correction schedule control function (option) Schedule management for periodically executing calibration correction configuration and the like.

An intuitive UI engineered for ease-of-use

Efficiently search for key data

Easily review historical data

Seamless display of historical trends—flick or drag the trend display to scroll through the data, even during measurement.

Easily review historical data

Quickly find data using calendars and summary screens

From a calendar, jump to waveforms of a specific date. From the alarm summary, jump to the waveform active during the alarm.

Quickly find data using calendars and summary screens

 

Easily check off trouble spots

Write freehand messages

Immediately clear areas of concern with a hand-written message.

Write freehand messages

You can draw or hand-write on the waveform area using a stylus (standard accessory) or the tip of your finger. You can even select a color and line width. Alternatively, you can select from a list of preset messages.

Save and output image files

Save trend waveforms of interest or screens displayed during alarms as image (PNG) files, and print them out at the same time.

Save and output image files

 

Check waveforms of concern in detail

Display digital values at any location

Move the scale to display the value corresponding to that position as a numeric value. Instantly check maximum/minimum measured values.

Display digital values at any location
[Patent technology]

Ascertain long-duration trends at a glance

All historical trends display

Long-duration trends can be fitted to a single screen for easy viewing.

All historical trends display
All historical trends display

Zoom in/out - time axis and engineering units

 

The time axis and engineering axis can expanded and compressed using a simple pinch together or apart function.

Pinch  apart / Pinch together
Pinch apart / Pinch together

 

Create your own screens

Custom display (/CG option)

You can arrange display objects such as trend, numeric, and bar graphs any way you like to create monitor displays that are customized to the environment. Start/stop pumps and perform other operations.

Custom display (/CG option)Custom display (/CG option)

Custom display building software
DAQStudio DXA170

DAQStudio is software for creating custom displays. You can load screens you created onto the GP via Ethernet or external memory media (SD/USB) and display them.

DAQStudio DXA170

 

Variety of display screens

Trend screen

Physical quantities are displayed and recorded on a log scale.

Physical quantities are displayed and recorded on a log scale.

Multi-panel display

You can select from 9 layouts, and save up to 20 configurations. (Multi panel available on the GP20 only)

Multi-panel displayMulti-panel display

Highly flexible and scalable architecture

Modular input/output

Inputs and outputs are modular for easy expandability. The GP multichannel paperless recorder main unit alone provides up to 100 channels (GP20) of measurement.

Modular input/output

 

Wide variety of input/output modules

Select from a wide variety of input/output modules.

Select from a wide variety of input /output modules.

Model Name Measurement/Application Channels
GX90XA-10-U2 Analog input module DC voltage, DC current (with external shunt resistor connected), thermocouple, RTD, contact (solid state relay scanner type) 10
GX90XA-10-L1 DC voltage, DC current (with external shunt resistor connected), thermocouple, contact (Low withstand voltage solid state relay scanner type) 10
GX90XA-10-T1 DC voltage, DC current (with external shunt resistor connected), thermocouple, contact (electromagnetic relay scanner type) 10
GX90XA-10-C1 DC current (mA) (solid state relay scanner type) 10
GX90XA-04-H0 DC voltage, DC current (with external shunt resistor connected), thermocouple, RTD, contact (individual A/D type) 4
GX90XA-06-R1 4-wire RTD, 4-wire resistance (solid state relay scanner type) 6
GX90YA Analog output module Current output 4
GX90XD Digital input module Remote control input or operation recording 16
GX90YD Digital output module Alarm output 6
GX90WD Digital input/output module Remote control input or operation recording/alarm output DI:8/DO:6
GX90XP Pulse input module Pulse signal data acquisition, integral count 10
GX90UT PID control module PID control (2 loop) AI:2/AO:2
DI:8/DO:8

Expandable to up to 450 channels (real actual input)

Supports up to 450 channels of measurement. Note that if MATH and communication channels are included, the GP20 large memory type can record on up to 1000 channels. The GP main unit and expandable I/O can both use the same input/output modules

Expandable to up to 450 channels (real actual input)

You connect directly with a LAN cable without connecting through a hub or repeater.
* You can also connect subunits of the GM Data Acquisition System.

Model Type Max.channels Number of channels by configuration
GP10 Standard 100ch Main unit only 0-30
Main + expandable I/O 0-100
GP20 Standard 100ch Main unit only 0-100
Main + expandable I/O 0-100
Large memory 450ch Main unit only 0-100
Main + expandable I/O 0-450

The number of channels is for analog input only.

Reduce wiring with distributed installation

When the recorder is installed offsite (away from the DUT), you can place the expandable I/O at the site and monitor data without the need for long-distance wiring of thermocouples and other sensors.

Reduce wiring with distributed installation

A full range of network functions and software

Real time remote monitoring from a web browser

Through a Web browser you can monitor the GP in real time and change settings. You can easily build a seamless, low-cost remote monitoring system with no additional software.

Real time monitoring screen

You can view monitor screens in real time that are identical to the trends, digital, and other displays on the GP main unit.

Real time monitoring screen (trend)

Enter settings online with a web browser

The setting screen lets you copy AI channel settings and other information to Excel for editing. You can reimport the data into the setting screen after editing.

Setting screen (AI channel)

Dedicated software (free download) is available for loading waveforms and GP settings.

Universal viewer

Data files saved on the GP can be viewed and printed. You can perform statistical computation over an area and export to ASCII, Excel, or other formats.

Universal viewer

Offline setting software

Save settings or transfer them to the GP.

Offline setting software

Aerospace Heat Treatment Supports heat treatment application AMS2750/NADCAP

Calibration correction schedule control function (/AH option)

Schedule management for periodically executing calibration correction configuration and the like. The correction factor can be set separately for unit and sensor dependency. TUS report software enables you to easily create TUS (temperature uniformity survey) reports.*.

* For information on TUS software, contact your Yokogawa representative.

Record data in separate files per equipment set

Multi-batch Function (/BT option)

Recorder pre-defined channel groups to separate data files with independent start and stop control. Up to 12 independent batches can be created.

PID control function

Control function

Enables PID and program control

  • PID control module
    2-loops per module, up to 20 loops per system
  • Setpoint program control function (/PG option)
    Up to 99 patterns
GX90UT PID control module

Remote operation and monitoring

The web application enables remote operation and monitoring from a browser.

Remote operation and monitoring

Built in control screens and display

Various pre-configured control screens and display are available.

Built in control screens and display

Seamless integration

Combine and integrate complex legacy control panel into a simple and flexible data acquisition station.

Simple instrumentation

Custom display

Remote operation is possible through screens that are customized for your specific system.

Custom display

High speed measurement (down to 1 ms)

Yokogawa's proprietary A/D converter allows the high speed module to measure data points as fast 1ms.

  • High speed (1 ms) measurement*
  • Proprietary A/D converter

* With 1ch per module. At 2 ms, 2 ch per module, and at 5 ms or more, all 4 ch per module.

GX90XA-04-H0 Analog input module (high speed AI)

Max. channels
Model Scan interval
1ms 2ms 10ms
GP10 1ch 5ch 10ch
GP20-1 1ch 5ch 10ch
GP20-2 5ch 25ch 40ch

Dual interval measurement with two different scan intervals

Users have the ability to choose two different scan intervals on a single GX/GP system. This allows users the flexibility to measure various types of inputs with two different scan intervals in a single system. For example, this provides for efficient, simultaneous measurement of signals with slow fluctuations such as temperature, and fast-changing signals such as pressure and vibration. Modules can be assigned to measurement groups.

Dual interval measurement with two different scan intervals

MATH function (/MT option)

Supports various kinds of math computation, including basic math and functions (square root, logarithms, trigonometry). Write formulas using variables for measured or computed data and save or display the results—this saves time and effort on post-processing. Create hourly, daily, monthly, and other reports with the Report function.

MATH function

Event actions

Ability to assign actions tied to specific events during the operation of the data acquisition station.

Event actions

Report creation and network functions (/MT option)

Report creation and network functions

Provides a variety of convenient networking functions

Modbus/TCP and Modbus/RTU Communications

GP supports Modbus TCP/IP client and server modes for Ethernet communications and Modbus RTU master and slave modes for optional serial communications.

Modbus/TCP (Ethernet connection)

Using the Modbus/TCP and Modbus/RTU functions, you can display and save data from the server and slave devices on the GP.*

* Requires the communication channel (/MC option ).

Modbus/TCP (Ethernet connection)

(Connect up to 16 Modbus/TCP servers, or up to 32 for the GP20-2.)
(Up to 31 Modbus/RTU slaves can be connected.)

Modbus RTU (RS-422A/485 connection)

Using the Modbus/TCP and Modbus/RTU functions, you can acquire GP data from upstream devices.

Modbus RTU (RS-422A/485 connection)

EtherNet/IP Function (/E1 option)

GP supports EtherNet/IP server functions. You can access GP from PLCs or other devices and load measurement/MATH channels or write* to communication input channels (GP10: max. 50 ch, GP20-1: max. 300 ch, GP20-2: max. 500 ch).

* Communication channel function (optional code, /MC) is required.

EtherNet/IP Function

CC-Link family SLMP communication (/E4 option)

Protocol function that enables connection from a GP to Mitsubishi Electric PLCs without sequencer programs. You can run the GP as an SLMP client, enabling writing of GP measured data to the PLC and writing of PLC data to communication channels.*

*Requires the communication channel function (/MC option).

SLMP Communication

Powerful tool for instrument performance evaluation testing (/E2 and /MC options)

Highly precise measured data from power measuring instruments (WT series power analyzers) can be acquired without loss of fidelity on the GP, and recorded and displayed alongside the GX's own measured data. This is ideal for performance evaluation testing because you can record instrument power consumption, temperature, and other phenomena simultaneously.

Powerful tool for instrument performance evaluation testing

  • Models that can be connected:
    Yokogawa Meters & Instruments Corp.,
    WT series power analyzers,
    WT300/WT300E (command mode WT300), WT500, WT1800/WT1800E (command type WT1800)
  • Max. no. of connections:
    8 (GP10), 16 (GP20)

 

OPC-UA Server (optional code /E3)

Data acquired by the GP can be accessed through Ethernet communication from a host system (OPCUA client). Writing from an upstream system to a GP communication channel requires the communication channel function (/MC option).

OPC-UA Server

The GX supports DARWIN communication commands. Use your current DARWIN communication programs as-is on the GP.*

* See your dealer or nearest Yokogawa representative for details.

DARWIN-compatible communication

CENTUM/STARDOM communciation package

  • CENTUM: LFS2432, DARWIN/DAQSTATION Communication package (for ALE111 [Ethernet])
  • STARDOM: NT365AJ DARWIN connection package

FTP-based file transfer

The FTP client/server functions allow you to easily share and manage data from a centralized file server.

FTP-based file transfer

 

E-mail messaging function

The GP can send a variety of informative e-mail messages that include alarm notification reports, periodic instantaneous data values, scheduled report data and other information.

E-mail messaging function

 

Automatic network setup (DHCP) function

Using Dynamic Host Configuration Protocol (DHCP), the GP can automatically acquire the settings it needs (IP address) for network communications from a DHCP server. This makes it easier than ever to install the unit on a plant network.

Automatic network setup (DHCP) function

Time synchronization with network time servers

GP uses SNTP protocol in client mode to acquire time information from a network time-server. This function allows any number of GP units within a facility to have precisely synchronized time; all units will record data with coordinated date and time stamp information. In addition, GP can function as a server, providing time data to other SNTP client units on the network.

Time synchronization with network time servers

Highly secure with proven reliability

Portable models

Our portable models are easy to take anywhere, and offer the same functionality and ease-of-use.

Portable models

Be confident that recorded data is saved

Measured and calculated data is continuously saved to secure, internal non-volatile memory. At manual or scheduled intervals, the files in memory are copied to the removable media. In addition, the files can be copied and archived to an FTP server.

Data redundancy

Because of the inherent reliability and security of non-volatile memory, the possibility of losing data under any operating condition or power failure event is extremely small.

 

Select file formats according to your application

For increased security, measured data can be saved in binary format. This format is very difficult to decipher or modify in traditional text editors or other programs. To enable easy and direct opening of the data in text editors or spreadsheet programs, choose text format. This allows you to work with your measurement data without dedicated software.

ASCI data display
ASCI data display

Binary data display
Binary data display

 

21 CFR Part 11 support (/AS option)

With the expanded security function option, the instruments support the USA FDA's Title 21 CFR Part 11 regulations (for the pharmaceutical manufacturing industry). It gives you access to a credential-based login function, electronic signatures, audit trails, an anti-tampering function, an Active Directory-based password management function, a sign-in function, and other security features.

21 CFR Part 11 support (/AS option)

Security enhancements

Safely sends and receives customer data.

>Security enhancements

SSL: An encryption protocol for data sent over TCP/IP networks.

Front panel door lock

The front panel door can be locked to prevent mishandling of the power switch or external media.

Front panel door lock

Analog front end module

A proprietary A/D converter delivers high speed, high precision data acquisition. (High-speed AI, PID control function)

Analog front end module

Standards supported

Standards supported

Analog input module

GX90XA

Specifications

Model GX90XA
Input type
(Inputs: 10)
DC voltage*1, standardized signal*1, thermocouple*1, RTD*2, DI*1, DC current (with external shunt resistor)*1, DC current*3, resistance*4
DCV 20 mV, 60 mV, 200 mV, 1 V, 2 V, 6 V, 20 V, 50 V, 100 V*5
Standard signal 0.4-2 V, 1-5 V
Resistance 20, 200, 2000 Ω
Thermocouple R, S, B, K, E, J, T, N, W, L, U, W97Re3-W75Re25, KpvsAu7Fe, Platinel 2, PR20-40, NiNiMo, W/WRe26, N(AWG14), XK GOST
RTD Pt100, JPt100, Cu10 GE, Cu10 L&N, Cu10 WEED, Cu10 BAILEY, Cu10 (20°C) α=0.00392, Cu10 (20°C) α=0.00393, Cu25 (0°C) α=0.00425, Cu53 (0°C) α=0.00426035, Cu100 (0°C) α=0.00425, J263B, Ni100 (SAMA), Ni100 (DIN), Ni120, Pt25, Pt50, Pt200 WEED, Cu10 GOST, Cu50 GOST, Cu100 GOST, Pt46 GOST, Pt100 GOST, PT500*4, PT1000*4
DI Level, Contact
DC current 0-20 mA, 4-20 mA
Scan intervals

1/2/5/10/20/50/100/200/500ms, 1/2/5s

Scan interval by type

Suffix code Scan interval
1ms 2ms 5ms 10ms 20ms 50ms 100ms 200ms 500ms 1s 2s 5s
-U2
-C1
-L1
-T1
-H0
-R1
Power supply and consumption Supplied from main unit, power consumption: 2 W or less
Insulation resistance Between input circuits and internal circuitry: 20 MΩ or greater (at 500 V DC)
Withstand voltage Between the input circuits and the internal circuitry: 3000 VAC for one minute (current input type and low withstand voltage type: 1500 VAC for one minute)
Between analog input channels: 1000 V AC for one minute (excluding b terminals for universal input type) (low withstand voltage type: 400 VAC for one minute, high speed universal type: 3000 V AC for one minute)
Terminal types M3 screw terminals or clamp terminals
Weight Approx. 0.3 kg

*1 Cannot be set for the current input type (type suffix code: -C1) or 4-wire RTD/resistance type (type suffix code: -R1).

*2 Cannot be set for the current input type (type suffix code: -C1), electromagnetic relay type (type suffix code: -T1), or low withstand voltage type (type suffix code: -L1).

*3 Can only be set with current input type (type suffix code: -C1).

*4 Can only be set with 4-wire RTD/resistance type (type suffix code: -R1).

*5 Can only be set with high speed universal type (type suffix code: -H0).

Model and Suffix Codes

Model Suffix Code Description
GX90XA           Analog input module
Number of channels -04         4 channels (-H0 type only)
-06         6 channels (-R1 type only)
-10         10 channels
Type -C1       Current, scanner type (isolated between channels)
-L1       Low withstand voltage DCV/TC/DI, scanner type (isolated between channels)
-U2       Universal, Solid state relay scanner type (3-wire RTD b-terminal common)
-T1       DCV/TC/DI, Electromagnetic relay scanner type (isolated between channels)
-H0       High speed universal, individual A/D type (isolated between channels)
-R1       4-wire RTD/resistance, scanner type (isolated between channels)
- N     Always N
Terminal form -3   Screw terminal (M3)
-C   Clamp terminal
Area N General

Analog output module

Specifications

Model GX90YA
Output type (outputs: 4) Transmission output, manual output
Range 4–20 mA or 0–20 mA
Output update interval 100 msec (shortest)
Load resistance 600 Ω or less
Resolution 0.002%
Power supply and consumption Supplied from main unit, power consumption: 3W or less
Insulation resistance Between output circuits and internal circuitry: 20 MΩ (at 500 VDC)
Between output channel terminals: 500 VDC, 20 MΩ or greater
Withstand voltage Between output circuits and internal circuitry: 1500 AC for one minute
Between output circuits: 500 VAC for one minute
Terminal type M3 screw terminals or clamp terminals
Weight Approximately 0.2 kg

Model and Suffix Codes

Model Suffix Code Description
GX90YA           Analog output module
Number of channels -04         4 channels
Type -C1       Current output (isolated between channels)
- N     Always N
Terminal form -3   Screw terminal (M3)
-C   Clamped terminals
Area N General

Digital input module

GX90XD

Specifications

Model GX90XD
Input types
(Inputs: 16)
  DI or pulse input*1 (Open collector or non-voltage contact)
ON/OFF detection Open collector: Voltage of 0.5 V DC or less when ON, leakage current of 0.5 mA or less when OFF
Non-voltage contact: Resistance of 200 Ω or less when ON, 50 kΩ when OFF
Contact rating 12 V DC, 20 mA or more
Power supply and consumption Supplied from main unit, power consumption: 0.7 W or less
Insulation resistance Between input terminals and internal circuitry: 20 MΩ or greater (at 500 V DC)
Withstand voltage Between input terminals and internal circuitry: 1500 V AC for one minute
Terminal types M3 screw terminals or clamp terminals
Weight Approx. 0.3 kg

Pulse input specifications*1

Counting system The rising edge of the pulse is counted.
Max. pulse period 250Hz (The chattering filter : OFF)
125Hz (The chattering filter : ON)
Minimum detection pulse width Low (close), High (open), both is 2 ms or more
Pulse detection period 1ms
Pulse measurement accuracy ± 1 pulse
Pulse count interval Measurement interval

Filter

The chattering filter can be switched On/Off. (When the chattering filter is off, connect GX/GP so that it is not affected by the noise.)

*1 Integration requires the math function (optional code /MT).

Model and Suffix Codes

Model Suffix Code Description
GX90XD           Digital input module
Number of channels -16         16 channels
Type -11       Open collector/Non-voltage, contact (shared common), Rated 5 VDC
- N     Always N
Terminal form -3   Screw terminal (M3)
-C   Clamp terminal
Area N General

Digital output module

GX90YD

Specifications

Model GX90YD
Output types (outputs: 6) Relay contact (c contact)
Rated load voltage 100 to 240 V AC or 5 to 24 V DC
Max. load voltage/current 264 VAC or 26.4 VDC, 3A/point (resistance load)
Power supply and consumption Supplied from main unit, power consumption: 1.4 W or less
Insulation resistance Between output terminals and internal circuitry: 20 MΩ (at 500 VDC)
Withstand voltage Between output terminals and internal circuitry: 3000 V AC for one minute
Terminal types M3 screw terminals
Weight Approx. 0.3 kg

Model and Suffix Codes

Model Suffix Code Description
GX90YD           Digital output module
Number of channels -06         6 channels
Type -11       Relay, SPDT(NO-C-NC)
- N     Always N
Terminal form -3   Screw terminal (M3)
Area N General

Expansion module

GX90EX

Model and Suffix Codes

Model Suffix Code Description
GX90EX          I/O expansion module
Port -02       2 Ports
Type -TP1     Twisted pair cable
- N   Always N
Area -N Standard accessories

Digital input/output module

GX90WD

Specifications

Model  GX90WD
Input types
(Inputs: 8)
  DI or pulse input*1 (Open collector or non-voltage contact)
ON/OFF detection Open collector : Voltage of 0.5 V DC or less when ON, leakage current of 0.5 mA or less when OFF
Non-voltage contact : Resistance of 200 Ω or less when ON, 50 kΩ when OFF
Contact input rating 12 VDC, 20 mA or more
Output type (outputs: 6)   Relay contact (C contact)
Rated load voltage When connected to the main circuit (fi rst-order power supply), 150 VAC or less
When connected to a circuit derived from the main circuit (second-order power supply), 250 VAC or less (the main circuit is 300 VAC or less and uses an isolated transformer) or 30 VDC or less
Max. load current 2 A (DC)/2 A (AC), resistive load
Power consumption 1.9 W or less
Insulation resistance Between input terminals and internal circuitry: 20 MΩ or greater (at 500 VDC)
Between output terminals and internal circuitry: 20 MΩ or greater (at 500 VDC)
Withstand voltage Between input terminals and internal circuitry: 1500 VAC for one minute
Between output terminals and internal circuitry: 3000 VAC for one minute
Terminal types M3 screw terminals
Weight Approx. 0.3 kg

Each unit (GX/GP main unit and expandable I/O), can use 1 module only.

Pulse input specifications*1

Counting system The rising edge of the pulse is counted.
Max. pulse period 250Hz (The chattering filter : OFF)
125Hz (The chattering filter : ON)
Minimum detection pulse width Low (close), High (open), both is 2 ms or more
Pulse detection period 1ms
Pulse measurement accuracy ± 1 pulse
Pulse count interval Measurement interval

Filter

The chattering filter can be switched On/Off. (When the chattering filter is off, connect GX/GP so that it is not affected by the noise.)

*1 Integration requires the math function (optional code /MT).

Model and Suffix Codes

Model Suffix Code Description
GX90WD           Digital input/output module
Number of channels -0806         8 channel DIs, 6 channel DOs
Type -01       Open collector/non-voltage contact (shared common), rated 5 VDC; Relay, SPDT, (NO-C-NC)
- N     Always N
Terminal form -3   Screw terminal (M3)
Area N General

Pulse input module

Specifications

Model GX90XP
Number of inputs 10
Measurement interval 100 ms (shortest)
Input type Contact (open collector, voltage-free contact), level (5 V logic)
Input range Up to 20 kHz*
* 30 Hz when the chattering filter is in use (On)
Minimum detection pulse width 25 μs*
* 15 ms when the chattering filter is in use (On)
Measurement accuracy Count ± 1 pulse
During integration, the following accuracies are added.
Upon MATH start: +1 measuring period
Upon MATH stop: -1 measuring period
* Integration requires the math function (optional code /MT).
Chattering filter Removes chattering up to 5 ms (can be turned on/off on each channel)
Hysteresis width Approx. 0.2 V
Contact, transistor rating Contact: 15 V DC or higher and 30 mA or higher rating.
Minimum applicable load current 1 mA or less.
Transistor: With the following ratings: Vce > 15 VDC, Ic > 30 mA
Maximum input voltage ±10 V DC
Insulation resistance Between input terminals and internal circuitry:
20 MΩ or greater at 500 V DC
Withstand voltage Between input terminals and internal circuitry: 1500 V AC for 1 minute

 

Model and Suffix Codes

Model Suffix code Description
GX90XP           Pulse input module
Number of channels -10         10 channels
Type -11       DC voltage/open collector/non-voltage contact (shared common), rated 5 VDC
N     Always N
Terminal form -3   Screw terminal (M3)
-C   Clamp terminal
Area N General

PID control module

Specifications

Model GX90UT
Control loops Number of loops 2
Analog input (measured input) Measured points 2
Measurement type DC voltage (DCV)/standardized signal, TC/RTD, DI (LEVEL and non-voltage contact)/DC current (with external shunt resistance)
Scan (control) interval 100 ms or 200 ms (system global setting)
Analog output (control output/transmission output/sensor power supply) Outputs 2
Output type Power supply for current, voltage pulse, or sensors.
Current output: 4-20 mA or 0-20 mA
Voltage pulse output: ON voltage = 12 VDC or more (load resistance 600 Ω or more), OFF voltage = 0.1 VDC or less
Can be used as a sensor power supply (13.0-18.3 VDC)
Digital input (switching the SP, operation mode, etc.) Inputs 8
Input format Non-voltage contact and open collector
Contact rating: 12 VDC or more, 20 mA or more
Digital output (of alarms, events, etc.) Outputs 8
Output format Open collector (sink type)
Output contact capacity Max 24 VDC, 50 mA
Withstand voltage/insulation resistance See PID control module general specifications (GS 04L51B01-31EN)
Terminal type M3 screw terminals
Weight Approximately 0.3kg

Model and Suffix Codes

Model Suffix Code Description
GX90UT           PID control module
Number of loops -02         2 loops
Type -11       8 DIs, 8 DOs
N     Always N
Terminal form -3   Screw terminals (M3)
Area N General

as of March, 2017

The SD cards of the following manufacturers have been verified for use with SMARTDAC+® GX/GP/GM, µR10000™/µR20000™ and FX1000™. However, please note that Yokogawa does not guarantee normal operation of the SD card when using with products listed below.

SMARTDAC+ GX
Applicable Models: GX10, GX20, GX20W
SMARTDAC+ GP
Applicable Models: GP10, GP20
SMARTDAC+ GM
Applicable Models: GM10

√: Verified
Manufacturer Capacity Manufacturer Model GX10 GX20 GP10 GP20 GX20W GM10
SanDisk 16GB SDSDXS-016G  
SanDisk   8GB SDSDXS-008G  
SanDisk 32GB SDSDXVE-032G
SanDisk 32GB SDSDXS-032G
SanDisk 32GB SDSDXL-032G
SanDisk 16GB SDSDXL-016G
SanDisk   8GB SDSDXL-008G

Applicable Models: µR10000/EM1, µR20000/EM1 

√: Verified
Manufacturer Capacity Manufacturer Model µR10000/EM1 µR20000/EM1
SanDisk 32GB SDSDXVE-032G
SanDisk 32GB SDSDXS-032G

Applicable Models: FX1000

√: Verified
Manufacturer Capacity Manufacturer Model FX1000
SanDisk 32GB SDSDXVE-032G
SanDisk 32GB SDSDXS-032G

In rare cases, the SD card is not recognized or generates an error when inserted into the instrument. If this occurrs, try the following.

  • Eject the SD card, then reinsert it.
  • Replace with a SD card that you know to be working normally.
  • Format the SD card on the Products (all data on the card will be deleted).

Please direct inquiries regarding the other manufacturers' verified SD cards above to their manufacturers.
Please note that sales of the above-mentioned SD card may be discontinued from the manufacturer without prior notice.

 

as of November, 2013

The USB barcode reader of the following manufacturers have been verified for use with the SMARTDAC+™ GX/GP series, DXAdvanced® series. However, please note that Yokogawa does not guarantee normal operation of the USB barcode reader when using with products listed below.
 

SMARTDAC+ GX/GP series
Applicable Models: GX10, GX20, GP10, GP20

DXAdvanced series
Applicable Models: DX1000、DX1000N、DX1000T、DX2000、DX2000T

Manufacturer Product Name
Honeywell Honeywell Voyager 9520 Series of Single-Line Laser Scanners
Honeywell Honeywell Xenon 1900 Area-Imaging Scanner

Please direct inquiries regarding the other manufacturers' verified USB barcode reader above to their manufacturers.
Please note that sales of the above-mentioned USB barcode reader may be discontinued from the manufacturer without prior notice.

Overview:

Temperature Uniformity Survey (TUS) is a vital part of the overall equipment performance and validation that ensures the heat treat/vacuum furnace repeatedly and accurately produces the same quality parts with no variance. Based upon the furnace class and its specifications, the equipment must hold uniform temperature across all zones at a specified set point and time-period, according to SAE AMS2750E pyrometry standards.

Overview:

Multi-touch technologies have rapidly moved from the commercial to the industrial sector where they are being used to enhance data analysis.

Brochures
Instruction Manuals
General Specifications
Software
Technical Information
Drawings

Product Overviews

    Overview:

    Smart user interface, smart architecture, and smart functionality is achieved in the new Yokogawa SMARTDAC+ data acquisition and control product series. The Yokogawa SMARTDAC+ GX and GP are fully integrated measurement, display, and recording platforms equipped with an advanced touch screen operator interface. GX series is a panel-mount design, capable of operating in harsh industrial applications and environments. GP is the portable version of the GX, intended for use in lab and test bench applications.

How-tos

Looking for more information on our people, technology and solutions?


Contact Us
Top