Errors caused by wet cables or bad connections.

It is very important that the cables maintain a very high impedance. If any moisture penetrates the cable the impedance will drop and the cable will short out the voltage generated by the pH sensors. This will cause errors or a total loss of measurement.  This can also occur if the cables are exposed to water for long periods of time. To insure that the pH measurement remains accurate protect the cables and the back of the pH sensors from exposure to moisture. 

Related Products & Solutions

  • pH and ORP Analyzers

    pH and ORP meters, analyzers and transmitters are used for continuous process monitoring of pH and ORP to ensure water/product quality, monitor effluent discharge, batch neutralization, pulp stock, scrubbers, cooling towers, chemical, water/wastewater treatment and many other applications.

    See More
  • pH and ORP Sensors

    • pH and ORP sensors
    • Proper pH electrode/sensor selection critical for optimal measurement results
    • Installation options include retractable, flow thru, immersion, and direct insertion
    See More