High Pressure/ High Temperature Conductivity Sensor SC42/SX42

High Pressure – High Temperature Contacting sensor primarily designed for Boiler Blowdown and Condensate Leak detection applications. These conductivity cells have extremely high temperature and pressure ratings: the threaded types can handle 16 bar at 200°C and the flanged types can handle 40 bar at 250°C.

Series (2)

   SC42 Small-Bore 2 SX42 Large Bore SX42 Large Bore 2
Model Code SC42 Small Bore SX42 Large Bore
Cell Constants 0.1/cm; 1.0/cm 0.1/cm; 1.0/cm 0.1/cm; 1.0/cm
Ranges 0.4-2,000 µS/cm; 4.0-5,000 µ/cm 0.02-2,000 µS/cm; 0.2-20,000 µS/cm 0.02-2,000 µS/cm; 0.2-20,000 µS/cm
Temperature Element Pt1000 Pt1000 Pt1000
Pressure Rating 7 bar/100 PSIG 16 bar/ 228 PSIG 40 bar/ 570 PSIG
Maximum Temperature 100°C/212°F up to 200°C up to 250°C
Installation Connections 3/4" NPT Fitting; 1", 1.5" or 2" Tri-Clamp 1" NPT 2" 600lb ANSI Flange
Integral Cable NO NO NO
Materials Of Construction 316 Stainless Steel 316 Stainless Steel 316 Stainless Steel
Design Sytle 2-Electrode 2-Electrode 2-Electrode
Overview:

Yes. The Cation Differential pH and ORP sensors have a glass electrode that needs a 2 point calibration performed regularly. Because of the Cation Reverence electrode you must use special buffers with a constant sodium ion concentration. This calibration is needed to compensate for the change in response of the glass electrodes. These changes are normal and are caused by aging or chemical attack, and they are accelerated at higher temperatures. 

Looking for more information on our people, technology and solutions?


Contact Us
Top